Yang Seung-Won, Kim Tai-Hyung, Choi Hyun-Min
Department of Physical Education, Baekseok University, Cheonan, Korea.
Graduate School of Technology Management, Kyung Hee University, Yongin, Korea.
J Exerc Rehabil. 2018 Aug 24;14(4):621-627. doi: 10.12965/jer.1836284.142. eCollection 2018 Aug.
The purpose of this study is to examine the validity and reproducibility of impedance body fat measurement devices measuring the body composition of Korean male and female adults using three bioelectrical impedance analyzers. We compared two methods for evaluating body composition: dual energy X-ray absorptiometry (DEXA), and bioelectrical impedance analysis (BIA). Subjects were 200 healthy adult Korean males and females whose mean±standard deviation (range) age, standing height, body weight, and body mass index (BMI) were 44.1±14.5 years, 172.8±7.4 cm, 76.0±12.8 kg, and 25.4±3.3 kg/m, and 44.5±14.7 years, 158.7±5.8 cm, 58.3±8.3 kg, and 23.2±3.0 kg/m, respectively. As a result, first of all, the reproducibility of the bioelectrical impedance analyzer had very high coefficients at =0.998, =0.997 between men and women, respectively. The correlation coefficients among three comparisons for lean body mass (LBM) were provided the following coefficients: =0.951 for DEXA vs. ACCUNIQ BC720, =0.950 for DEXA vs. ACCUNIQ BC360, and =0.946 for DEXA vs. ACCUNIQ BC380 in men. In the results for women, they also had the very high following coefficients: =0.956 for DEXA vs. ACCUNIQ BC720, =0.946 for DEXA vs. ACCUNIQ BC360, and =0.957 for DEXA vs. ACCUNIQ BC380 in LBM. In conclusion, this research showed a higher correlation in terms of accuracy compared to existing BIA-based body composition measurement techniques, and the accuracy of LBM was improved with high correlation coefficients through the algorithm that was improved using the multifrequency BIA method in the ACCUNIQ BC products.
本研究的目的是使用三种生物电阻抗分析仪,检验测量韩国成年男性和女性身体成分的阻抗体脂测量设备的有效性和可重复性。我们比较了两种评估身体成分的方法:双能X线吸收法(DEXA)和生物电阻抗分析(BIA)。研究对象为200名健康的韩国成年男性和女性,其平均±标准差(范围)年龄、身高、体重和体重指数(BMI)分别为44.1±14.5岁、172.8±7.4厘米、76.0±12.8千克和25.4±3.3千克/平方米,以及44.5±14.7岁、158.7±5.8厘米、58.3±8.3千克和23.2±3.0千克/平方米。结果,首先,生物电阻抗分析仪的可重复性在男性和女性中分别具有非常高的系数,男性为r = 0.998,女性为r = 0.997。在三种比较中,去脂体重(LBM)的相关系数如下:男性中,DEXA与ACCUNIQ BC720相比r = 0.951,DEXA与ACCUNIQ BC360相比r = 0.950,DEXA与ACCUNIQ BC380相比r = 0.946。在女性结果中,她们在LBM方面也有非常高的系数:DEXA与ACCUNIQ BC720相比r = 0.956,DEXA与ACCUNIQ BC360相比r = 0.946,DEXA与ACCUNIQ BC380相比r = 0.957。总之,本研究表明,与现有的基于BIA的身体成分测量技术相比,在准确性方面具有更高的相关性,并且通过在ACCUNIQ BC产品中使用多频BIA方法改进的算法,LBM的准确性得到提高,相关系数很高。