依赖于MRI扫描参数的纹理分析特征的定量变化:一个体模模型。

Quantitative variations in texture analysis features dependent on MRI scanning parameters: A phantom model.

作者信息

Buch Karen, Kuno Hirofumi, Qureshi Muhammad M, Li Baojun, Sakai Osamu

机构信息

Department of Radiology, Boston Medical Center, Boston University School of Medicine, Boston, MA, USA.

Department of Diagnostic Radiology, National Cancer Center Hospital East, Kashiwa, Chiba, Japan.

出版信息

J Appl Clin Med Phys. 2018 Nov;19(6):253-264. doi: 10.1002/acm2.12482. Epub 2018 Oct 27.

Abstract

OBJECTIVES

To evaluate the influence of MRI scanning parameters on texture analysis features.

METHODS

Publicly available data from the Reference Image Database to Evaluate Therapy Response (RIDER) project sponsored by The Cancer Imaging Archive included MRIs on a phantom comprised of 18 25-mm doped, gel-filled tubes, and 1 20-mm tube containing 0.25 mM Gd-DTPA (EuroSpinII Test Object5, Diagnostic Sonar, Ltd, West Lothian, Scotland). MRIs performed on a 1.5 T GE HD, 1.5 T Siemens Espree (VB13), or 3.0 T GE HD with TwinSpeed gradients with an eight-channel head coil included T1WIs with multiple flip angles (flip-angle = 2,5,10,15,20,25,30), TR/TE = 4.09-5.47/0.90-1.35 ms, NEX = 1 and DCE with 30° flip-angle, TR/TE=4.09-5.47/0.90-1.35, and NEX = 1,4. DICOM data were imported into an in-house developed texture analysis program which extracted 41-texture features including histogram, gray-level co-occurrence matrix (GLCM), and gray-level run-length (GLRL). Two-tailed t tests, corrected for multiple comparisons (Q values) were calculated to compare changes in texture features with variations in MRI scanning parameters (magnet strength, flip-angle, number of excitations (NEX), scanner platform).

RESULTS

Significant differences were seen in histogram features (mean, median, standard deviation, range) with variations in NEX (Q = 0.003-0.045) and scanner platform (Q < 0.0001), GLCM features (entropy, contrast, energy, and homogeneity) with NEX (Q = 0.001-0.018) and scanner platform (Q < 0.0001), GLRL features (long-run emphasis, high gray-level run emphasis, high gray-level emphasis) with magnet strength (Q = 0.0003), NEX (Q = 0.003-0.022) and scanner platform (Q < 0.0001).

CONCLUSION

Significant differences were seen in many texture features with variations in MRI acquisition emphasizing the need for standardized MRI technique.

摘要

目的

评估MRI扫描参数对纹理分析特征的影响。

方法

来自由癌症影像存档库赞助的评估治疗反应参考图像数据库(RIDER)项目的公开数据,包括对一个体模的MRI扫描,该体模由18根25毫米掺杂的凝胶填充管和1根20毫米管组成,后者含有0.25 mM的钆喷酸葡胺(EuroSpinII测试物体5,Diagnostic Sonar有限公司,西洛锡安,苏格兰)。在1.5T GE HD、1.5T西门子Espree(VB13)或3.0T GE HD且带有双速梯度和八通道头部线圈的设备上进行的MRI扫描,包括具有多个翻转角(翻转角 = 2、5、10、15、20、25、30)的T1加权成像,TR/TE = 4.09 - 5.47/0.90 - 1.35毫秒,激励次数(NEX)= 1,以及具有30°翻转角、TR/TE = 4.09 - 5.47/0.90 - 1.35且NEX = 1、4的动态对比增强成像。DICOM数据被导入到一个内部开发的纹理分析程序中,该程序提取了41个纹理特征,包括直方图、灰度共生矩阵(GLCM)和灰度游程长度(GLRL)。计算经过多重比较校正(Q值)的双尾t检验,以比较纹理特征的变化与MRI扫描参数(磁场强度、翻转角、激励次数(NEX)和扫描平台)的变化。

结果

在直方图特征(均值、中位数、标准差、范围)方面,随着NEX(Q = 0.003 - 0.045)和扫描平台(Q < 0.0001)的变化存在显著差异;在GLCM特征(熵、对比度、能量和均匀性)方面,随着NEX(Q = 0.001 - 0.018)和扫描平台(Q < 0.0001)的变化存在显著差异;在GLRL特征(长游程强调、高灰度游程强调、高灰度强调)方面,随着磁场强度(Q = 0.0003)、NEX(Q = 0.003 - 0.022)和扫描平台(Q < 0.0001)的变化存在显著差异。

结论

随着MRI采集参数的变化,许多纹理特征存在显著差异,这强调了标准化MRI技术的必要性。