Cataract quantification using swept-source optical coherence tomography.

机构信息

From the SciTec Department (Brás, Sickenberger), JenVis Research GbR, Ernst-Abbe-Hochschule Jena, Jena, Germany Vienna Institute for Research in Ocular Surgery (Hirnschall, Findl), Ophthalmology Department, Hanusch-Hospital, Vienna, Austria.

From the SciTec Department (Brás, Sickenberger), JenVis Research GbR, Ernst-Abbe-Hochschule Jena, Jena, Germany Vienna Institute for Research in Ocular Surgery (Hirnschall, Findl), Ophthalmology Department, Hanusch-Hospital, Vienna, Austria.

出版信息

J Cataract Refract Surg. 2018 Dec;44(12):1478-1481. doi: 10.1016/j.jcrs.2018.08.009. Epub 2018 Nov 2.

Abstract

PURPOSE

To develop and evaluate a cataract quantification method using a swept-source optical coherence tomography (SS-OCT) device (IOLMaster 700).

SETTING

Hanusch Hospital, Vienna, Austria.

DESIGN

Prospective multicenter case series.

METHODS

This study included patients with cataract in at least 1 eye. Two independent examiners performed Lens Opacities Classification System II (LOCS II) grading at the slitlamp independently. Corrected distance visual acuity, contrast sensitivity, and SS-OCT measurements were also performed. In addition, the phacoemulsification energy and time were recorded. To develop an objective SS-OCT-based cataract quantification system, all SS-OCT scans were segmented and the local pixel intensity unit of the lens was analyzed using Matlab's grayscale counting. The pixel intensity unit of the lens nucleus (OCT-based cataract quantification system score) was equated to the clinician's subjective nuclear opalescence grading.

RESULTS

The study evaluated 186 eyes (113 patients). The correlation between the independent examiners' LOCS grading was good (0.91) (P < .01). The correlation between the LOCS grading and OCT-based cataract quantification system score was 0.86 for examiner 1 and 0.76 for examiner 2 (both P < .01). In 24 patients, phacoemulsification time, power, and energy; visual acuity; and contrast sensitivity were available and included in the study. The OCT-based cataract quantification system scores correlated significantly with phacoemulsification time (0.71) and energy (0.64) (both P < .01).

CONCLUSIONS

Cataract density could be evaluated using an SS-OCT device, and the OCT-based cataract quantification system values correlated with the conventional LOCS II classification. Swept-source OCT measurements allowed quantification and documentation cataract density.

摘要

目的

开发和评估一种使用扫频源光学相干断层扫描(SS-OCT)设备(IOLMaster 700)的白内障定量方法。

设置

奥地利维也纳 Hanusch 医院。

设计

前瞻性多中心病例系列。

方法

本研究纳入了至少 1 只眼患有白内障的患者。2 位独立的检查者在裂隙灯下独立进行 Lens Opacities Classification System II(LOCS II)分级。还进行了矫正距离视力、对比敏感度和 SS-OCT 测量。此外,还记录了白内障超声乳化的能量和时间。为了开发一种基于 SS-OCT 的白内障定量系统,对所有 SS-OCT 扫描进行了分割,并使用 Matlab 的灰度计数分析晶状体的局部像素强度单位。晶状体核的像素强度单位(基于 OCT 的白内障定量系统评分)与临床医生的主观核混浊分级相当。

结果

该研究评估了 186 只眼(113 例患者)。2 位独立检查者的 LOCS 分级相关性良好(0.91)(P<.01)。LOCS 分级与基于 OCT 的白内障定量系统评分的相关性为检查者 1 为 0.86,检查者 2 为 0.76(均 P<.01)。在 24 例患者中,可获得白内障超声乳化时间、功率和能量、视力和对比敏感度,并将其纳入研究。基于 OCT 的白内障定量系统评分与白内障超声乳化时间(0.71)和能量(0.64)显著相关(均 P<.01)。

结论

可以使用 SS-OCT 设备评估白内障密度,基于 OCT 的白内障定量系统值与传统 LOCS II 分类相关。扫频 OCT 测量允许定量和记录白内障密度。