Average lens density quantification with swept-source optical coherence tomography: optimized, automated cataract grading technique.

机构信息

Department of Ophthalmology, Rothschild Foundation, Paris, France.

Department of Ophthalmology, Rothschild Foundation, Paris, France; American University of Beirut, Lebanon.

出版信息

J Cataract Refract Surg. 2019 Dec;45(12):1746-1752. doi: 10.1016/j.jcrs.2019.07.033.

Abstract

PURPOSE

To assess a revised version of an objective cataract grading method based on the lens densitometry on swept-source optical coherence tomography (SS-OCT) scans.

SETTING

Rothschild Foundation, Paris, France.

DESIGN

Prospective case series.

METHODS

All patients seeking a cataract evaluation and consenting to study participation were included. Exclusion criteria were a history of ocular surgery, cornea or retina disorders, and ocular dryness. The average lens densitometry (ALD) was measured on SS-OCT scans (IOLMaster 700) using an algorithm developed with MATLAB software. The ocular scatter index (OSI), measured using a double-pass aberrometer (Optical Quality Analysis System); Pentacam nucleus staging (PNS) score; and mean PNS were also measured and compared with the ALD measurements.

RESULTS

Two hundred eighty-five eyes (195 with cataract; 90 controls) of 155 patients aged 20 to 93 years (mean 56.5 years ± 21 [SD]) were included. The ALD measurements were repeatable (P = .99, analysis of variance). The repeatability limit was 2.50 pixel units. The ALD correlated with the OSI (r = 0.41, P < .01) and PNS score (r = 0.65, P < .01). An ALD greater than 73.8 pixel units was the cutoff threshold for cataract, with 96.2% sensitivity and 91.3% specificity. Discriminant analysis produced a model combining visual acuity and the mean ALD with an area under the receiver operating characteristic curve of 0.975 (P < .001), where the mean ALD had the highest coefficient.

CONCLUSIONS

Using the automated algorithm based on SS-OCT ALD measurements was a repeatable, reliable objective cataract grading method. When the ALD is greater than 73.8 pixel units, it is reasonable to discuss cataract surgery in symptomatic patients.

摘要

目的

评估基于扫频源光学相干断层扫描(SS-OCT)扫描的晶状体密度计的改良客观白内障分级方法。

设置

法国巴黎罗斯柴尔德基金会。

设计

前瞻性病例系列。

方法

所有寻求白内障评估并同意参与研究的患者均被纳入。排除标准为眼部手术史、角膜或视网膜疾病以及眼干燥症。使用 MATLAB 软件开发的算法在 SS-OCT 扫描(IOLMaster 700)上测量平均晶状体密度(ALD)。使用双光折射仪(光学质量分析系统)测量眼散射指数(OSI);Pentacam 核分期(PNS)评分;并与 ALD 测量值进行比较。

结果

155 名患者的 285 只眼(195 只白内障眼;90 只对照眼)年龄为 20 至 93 岁(平均 56.5 ± 21 [SD] 岁)。ALD 测量值具有可重复性(P =.99,方差分析)。重复性限为 2.50 像素单位。ALD 与 OSI(r = 0.41,P <.01)和 PNS 评分(r = 0.65,P <.01)相关。ALD 大于 73.8 像素单位是白内障的截断阈值,灵敏度为 96.2%,特异性为 91.3%。判别分析生成了一个结合视力和平均 ALD 的模型,受试者工作特征曲线下面积为 0.975(P <.001),其中平均 ALD 的系数最高。

结论

使用基于 SS-OCT ALD 测量的自动算法是一种可重复、可靠的客观白内障分级方法。当 ALD 大于 73.8 像素单位时,对有症状的患者进行白内障手术是合理的。