Analysis of Goldman Applanation Tonometry with and without fluorescein among glaucomatous and nonglaucomatous patients attending Mzuzu Central Hospital in Malawi: A cross-sectional study.

作者信息

Munyapa Tryness, Mzumara Thokozani, Ogbonna Grace, Mvula Augustine

机构信息

Department of Optometry Mzuzu University Mzuzu Malawi.

Department of Ophthalmology Mzimba North District Hospital Mzuzu Malawi.

出版信息

Health Sci Rep. 2023 Jul 1;6(7):e1390. doi: 10.1002/hsr2.1390. eCollection 2023 Jul.

Abstract

BACKGROUND AND AIM

Goldman Applanation Tonometry (GAT), the gold standard of tonometry, is used without fluorescein in low-resource settings. Nevertheless, corneal biomechanics differ among population groups.

PURPOSE

The aim of the study is to assess the relationship between GAT findings with and without fluorescein among glaucomatous and nonglaucomatous adults in Malawi.

METHODS

This was a cross-sectional quantitative study involving 22 glaucoma patients and 22 nonglaucoma patients at Mzuzu Central Hospital. We used a purposive sampling technique to select participants into the two groups. Next, we measured intraocular pressure using GAT with and without fluorescein. Then we entered the data into SPSS version 25. We employed the Wilcoxon test to make comparisons based on age and gender. We considered the value of  < 0.05 statistically significant.

RESULTS

There is a statistically significant and strong positive correlation between nfGAT and fGAT among both glaucoma ( = 0.989,  < 0.001) and nonglaucoma ( = 0.955,  < 0.001). According to age, there is no significant difference in IOP value measured with nfGAT and fGAT for both glaucomas ( = 0.109) and nonglaucoma subjects ( = 0.076). However, significant differences were observed between nfGAT and fGAT mean IOP according to sex among both glaucomatous and nonglaucomatous subjects ( = 0.017 and  = 0.32, respectively).

CONCLUSION

The study suggests that the merits of intraocular pressure measured using GAT without fluorescein are not speculative, therefore the two techniques can be routinely used interchangeably in diagnosing and managing glaucoma.

摘要

背景与目的

戈德曼压平眼压计(GAT)是眼压测量的金标准,在资源匮乏地区使用时无需荧光素。然而,不同人群的角膜生物力学存在差异。

目的

本研究旨在评估马拉维青光眼和非青光眼成年人中使用荧光素与不使用荧光素时GAT测量结果之间的关系。

方法

这是一项横断面定量研究,在姆祖祖中心医院纳入了22例青光眼患者和22例非青光眼患者。我们采用目的抽样技术将参与者分为两组。接下来,我们使用GAT分别在有荧光素和无荧光素的情况下测量眼压。然后将数据录入SPSS 25版。我们采用威尔科克森检验根据年龄和性别进行比较。我们认为P < 0.05具有统计学意义。

结果

在青光眼患者(r = 0.989,P < 0.001)和非青光眼患者(r = 0.955,P < 0.001)中,非荧光素GAT(nfGAT)和荧光素GAT(fGAT)之间均存在统计学显著且强的正相关。根据年龄,青光眼患者(P = 0.109)和非青光眼受试者(P = 0.076)使用nfGAT和fGAT测量的眼压值无显著差异。然而,在青光眼和非青光眼受试者中,根据性别观察到nfGAT和fGAT平均眼压之间存在显著差异(分别为P = 0.017和P = 0.32)。

结论

该研究表明,不使用荧光素的GAT测量眼压的优点并非臆测,因此这两种技术在青光眼的诊断和管理中可常规交替使用。