1 Department of Cardiology University of Chicago IL.
2 Cardiovascular Research Foundation and Columbia University New York NY.
J Am Heart Assoc. 2018 Nov 20;7(22):e009175. doi: 10.1161/JAHA.118.009175.
Background Right heart catheterization is the gold standard in clinical practice for the assessment of cardiovascular hemodynamics, but it is an invasive procedure requiring expertise in both insertion and reading. Remote dielectric sensing (Re DS ) is a noninvasive electromagnetic-based technology intended to quantify lung fluid content. Methods and Results In this prospective single-center study, Re DS readings were obtained in supine position just before right heart catheterization procedure in patients with heart failure. Agreement between Re DS and pulmonary artery wedge pressure ( PAWP ) was analyzed. Of all, 139 patients with heart failure received hemodynamic assessment and Re DS measurement. A good correlation was found between Re DS and PAWP measurement ( r=0.492, P<0.001). Receiver operating characteristic analysis of the ability to identify a PAWP ≥18 mm Hg resulted in a Re DS cutoff value of 34%, with an area under the curve of 0.848, a sensitivity of 90.7%, and a specificity of 77.1%. Overall, Re DS <34% carries a high negative predictive value of 94.9%. Conclusions Lung fluid content, as measured by Re DS , correlates well with PAWP . The high sensitivity and specificity and especially the high negative predictive value make Re DS a reliable noninvasive tool at the point of care, to rule out elevated PAWP in patients with heart failure and to help with medical management of patients with heart failure. Further studies are warranted to compare this tool with existing tests and to relate the findings to the clinical outcomes.
右心导管检查是评估心血管血流动力学的临床金标准,但它是一种需要在插入和读取方面都有专业知识的侵入性操作。远程介电感应(Re DS)是一种非侵入性的基于电磁的技术,旨在量化肺液含量。
在这项前瞻性单中心研究中,在心力衰竭患者接受右心导管检查前的仰卧位获得 Re DS 读数。分析了 Re DS 与肺动脉楔压(PAWP)之间的一致性。共有 139 例心力衰竭患者接受了血流动力学评估和 Re DS 测量。Re DS 与 PAWP 测量之间存在良好的相关性(r=0.492,P<0.001)。PAWP≥18mmHg 的 Re DS 识别能力的接收者操作特征分析得出 Re DS 截断值为 34%,曲线下面积为 0.848,灵敏度为 90.7%,特异性为 77.1%。总体而言,Re DS<34%具有 94.9%的高阴性预测值。
Re DS 测量的肺液含量与 PAWP 相关性良好。高灵敏度、特异性,尤其是高阴性预测值使 Re DS 成为心力衰竭患者床边可靠的非侵入性工具,可以排除 PAWP 升高,并有助于心力衰竭患者的医疗管理。需要进一步的研究来比较这种工具与现有测试,并将结果与临床结果相关联。