Am J Ophthalmol. 2019 Apr;200:255-262. doi: 10.1016/j.ajo.2019.01.015. Epub 2019 Jan 26.
To evaluate intra-subject repeatability of anterior segment optical coherence tomography (AS-OCT) combined with placido disc MS-39 (CSO, Firenze, Italy) and correlate epithelial thickness measurements and the degree of visual limitation in keratoconus patients.
Reliability analysis and cross-sectional study.
Setting: Vissum, Alicante, Spain.
Total of 167 eyes, of which 107 were from keratoconus patients and 60 were healthy.
Three repeated consecutive epithelial thickness measurements were taken: central 3.0 mm and the nasal, temporal, superior, and inferior regions at 3.0-6.0 mm and 6.0-8.0 mm.
Intrasubject repeatability of epithelial indices was assessed and correlated to the degree of best-corrected visual acuity (BCVA).
The intraclass correlation coefficient was ≥0.90, indicating highly repeatable measurements for both keratoconus and healthy eyes. In multivariate analysis, epithelial indices associated with worse BCVA were as follows: thinner 3 mm central (P = .04), thicker 8 mm superior (P < .001), and thinner 8 mm inferior (P < .001). Thickness central 3 mm decreased as keratoconus grading increased (P = .002). The superior-inferior ratio at 8 mm significantly increased as keratoconus grading increased (P < .001). In multivariate analysis thinner 3 mm central (R = 6.19%, P = .04), greater superior-inferior 8 mm ratio (R = 8.37%, P = .004), greater superior-inferior 6mm ratio (R = 2.67%, P = .02), and steeper K2 (R = 21.51%, P < .001) were associated with keratoconus with a combined area under the curve of 0.92.
High repeatability of epithelial thickness measurements may be achieved in both keratoconus and healthy eyes with the MS-39. There is a significant correlation between epithelial thickness measurements and BCVA. These measurements may be of additional value in discriminating between keratoconus and healthy eyes.
评估眼前节光学相干断层扫描(AS-OCT)联合角膜地形图 MS-39(CSO,意大利佛罗伦萨)的同患者内重复性,并与圆锥角膜患者的上皮厚度测量值和视觉受限程度相关联。
可靠性分析和横断面研究。
西班牙阿利坎特的 Vissum。
共 167 只眼,其中 107 只为圆锥角膜患者,60 只为健康对照。
连续 3 次测量上皮厚度:中央 3.0mm 和鼻侧、颞侧、上侧和下侧 3.0-6.0mm 和 6.0-8.0mm 区域。
评估上皮指数的同患者内重复性,并与最佳矫正视力(BCVA)的程度相关联。
内类相关系数≥0.90,表明圆锥角膜和健康眼的测量值高度可重复。多元分析显示,与 BCVA 较差相关的上皮指数如下:中央 3mm 处更薄(P=0.04)、上侧 8mm 处更厚(P<0.001)和下侧 8mm 处更薄(P<0.001)。随着圆锥角膜分级的增加,中央 3mm 厚度减小(P=0.002)。随着圆锥角膜分级的增加,8mm 处的上下比显著增加(P<0.001)。多元分析显示,中央 3mm 更薄(R=6.19%,P=0.04)、上下 8mm 比更大(R=8.37%,P=0.004)、上下 6mm 比更大(R=2.67%,P=0.02)和 K2 更陡峭(R=21.51%,P<0.001)与圆锥角膜相关,曲线下面积为 0.92。
角膜地形图 MS-39 可在圆锥角膜和健康眼中实现上皮厚度测量的高度可重复性。上皮厚度测量值与 BCVA 之间存在显著相关性。这些测量值在区分圆锥角膜和健康眼方面可能具有额外的价值。