Carleton University, 1125 Colonel By Dr, Ottawa, ON, K1S 5B6, Canada.
University of Ottawa Heart Institute, 40 Ruskin St, Ottawa, ON, K1Y 4W7, Canada.
Med Phys. 2019 Apr;46(4):1697-1706. doi: 10.1002/mp.13419. Epub 2019 Feb 18.
Patient body motion is known to cause large deviations in the determination of myocardial blood flow (MBF) with errors exceeding 300%. Accurate correction for patient whole-body motion is still a largely unsolved problem in cardiac positron emission tomography (PET) imaging.
This study evaluated the efficacy of using Natterer's formulation of the Helgason-Ludwig consistency conditions on the two-dimensional Radon transform to align computed tomography to PET projection data in multiple time frames of a dynamic sequence for the purpose of frame-by-frame correction of rigid whole-body motion.
The correction algorithm was evaluated with digital NCAT phantoms using realistic noise added by the analytical simulator. Count rates used in the simulation were derived from clinical patient data. In addition, a proof of concept test using measured data with a cardiac torso phantom was conducted.
Motion correction resulted in significant improvement in the accuracy of MBF estimates, especially for high count-rate acquisitions. Maximum errors for 2 cm of motion dropped from 325% to 25% and from 250% to 25% using global and regional partial-volume correction, respectively. Median MBF errors dropped from 33% to 4.5% and 27% to 3.8%, respectively. Importantly, the correction algorithm performed equally well to compensate for body motion in both early and late time frames.
Cardiac PET-CT data used for attenuation correction (CTAC) alignment using projection consistency conditions was effective for reducing errors in MBF measurements due to simulated patient motion, and can be integrated into the image reconstruction workflow.
众所周知,患者身体运动会导致心肌血流(MBF)的测定产生较大偏差,误差超过 300%。准确校正患者全身运动仍然是心脏正电子发射断层扫描(PET)成像中的一个尚未解决的大问题。
本研究评估了使用 Natterer 的 Helgason-Ludwig 一致性条件在二维 Radon 变换上的公式对齐计算断层摄影术与 PET 投影数据在动态序列的多个时间帧中,以便对刚性全身运动进行逐帧校正的功效。
使用具有分析模拟器添加的真实噪声的数字 NCAT 体模评估校正算法。模拟中使用的计数率是从临床患者数据中得出的。此外,还使用心脏体模的测量数据进行了概念验证测试。
运动校正导致 MBF 估计的准确性有了显著提高,尤其是在高计数率采集时。使用全局和区域部分容积校正,运动 2 厘米的最大误差分别从 325%降至 25%和从 250%降至 25%。中位数 MBF 误差分别从 33%降至 4.5%和从 27%降至 3.8%。重要的是,校正算法在早期和晚期时间帧中都能很好地补偿身体运动。
用于衰减校正(CTAC)的心脏 PET-CT 数据使用投影一致性条件对齐对于减少因模拟患者运动引起的 MBF 测量误差是有效的,并且可以集成到图像重建工作流程中。