Department of Radiological and Medical Laboratory Sciences, Nagoya University Graduate School of Medicine.
Brain & Mind Research Center, Nagoya University.
Magn Reson Med Sci. 2019 Oct 15;18(4):265-271. doi: 10.2463/mrms.mp.2018-0063. Epub 2019 Mar 4.
The accuracy of flow velocity and three-directional velocity components are important for the precise visualization of hemodynamics by 3D cine phase-contrast MRI (3D cine PC MRI, also referred to as 4D-flow). The aim of this study was to verify the accuracy of these measurements of prototype or commercially available 3D cine PC MRI obtained by three different manufactures' MR scanners.
The verification of the accuracy of flow velocity in 3D cine PC MRI was performed by circulating blood mimicking fluid through a straight-tube phantom in a slanting position, such that the three-directional velocity components were simultaneously measurable, using three 3T MR scanners from different manufacturers. The data obtained were processed by phase correction, and the velocity and three-directional velocity components in the center of the tube on the central cross section of a slab were calculated. The velocity profile in each three directions and the composite velocity profiles were compared with the calculated reference values, using the Hagen-Poiseuille equation. In addition, velocity profiles and the spatially time-averaged velocity perpendicular to the tube were compared with the theoretical values and measured values by a flowmeter, respectively.
An underestimation of the maximum velocity in the center of the tube and an overestimation of the velocity near the tube wall due to partial volume effects were observed in all three scanners. A roughening and flattening of profiles in the center of the tube were observed in one scanner, due, presumably, to the low signal-to-noise ratio. However, the spatially time-averaged velocities corresponded well with the measured values by the flowmeter in all three scanners.
In this study, we have demonstrated that the accuracy of flow velocity and three-directional velocity components in 3D cine PC MRI was satisfactory in all three MR scanners.
在 3D 电影相位对比 MRI(3D 电影 PC MRI,也称为 4D-flow)中,流速和三维速度分量的准确性对于精确显示血液动力学非常重要。本研究的目的是验证由三个不同制造商的磁共振扫描仪获得的原型或市售 3D 电影 PC MRI 的这些测量值的准确性。
通过在倾斜位置的直管体模中循环血液模拟液,同时测量三个方向的速度分量,在三个来自不同制造商的 3T 磁共振扫描仪上验证 3D 电影 PC MRI 中的流速准确性。使用相位校正处理获得的数据,并计算管中心在板的中心横截面上的速度和三维速度分量。使用 Hagen-Poiseuille 方程将每个三个方向的速度分布和复合速度分布与计算的参考值进行比较。此外,分别将速度分布和垂直于管的空间时间平均速度与理论值和流量计测量值进行比较。
在所有三个扫描仪中,都观察到由于部分体积效应,管中心的最大速度低估和管壁附近的速度高估。在一个扫描仪中,管中心的轮廓变得粗糙和平坦,可能是由于信噪比低所致。然而,在所有三个扫描仪中,空间时间平均速度都与流量计测量值非常吻合。
在这项研究中,我们证明了所有三个磁共振扫描仪中 3D 电影 PC MRI 的流速和三维速度分量的准确性令人满意。