Suppr超能文献

在一项多软件和多供应商交叉对比研究中,各种 4D Flow MRI 后处理软件程序的重复性和可再现性。

Repeatability and reproducibility of various 4D Flow MRI postprocessing software programs in a multi-software and multi-vendor cross-over comparison study.

机构信息

Department of Radiology and Nuclear Medicine, Universität zu Lübeck, Ratzeburger Allee 160, 23538, Lübeck, Germany.

Center of Brain, Behavior and Metabolism (CBBM), Universität zu Lübeck, Lübeck, Germany.

出版信息

J Cardiovasc Magn Reson. 2023 Mar 28;25(1):22. doi: 10.1186/s12968-023-00921-4.

Abstract

BACKGROUND

Different software programs are available for the evaluation of 4D Flow cardiovascular magnetic resonance (CMR). A good agreement of the results between programs is a prerequisite for the acceptance of the method. Therefore, the goal was to compare quantitative results from a cross-over comparison in individuals examined on two scanners of different vendors analyzed with four postprocessing software packages.

METHODS

Eight healthy subjects (27 ± 3 years, 3 women) were each examined on two 3T CMR systems (Ingenia, Philips Healthcare; MAGNETOM Skyra, Siemens Healthineers) with a standardized 4D Flow CMR sequence. Six manually placed aortic contours were evaluated with Caas (Pie Medical Imaging, SW-A), cvi42 (Circle Cardiovascular Imaging, SW-B), GTFlow (GyroTools, SW-C), and MevisFlow (Fraunhofer Institute MEVIS, SW-D) to analyze seven clinically used parameters including stroke volume, peak flow, peak velocity, and area as well as typically scientifically used wall shear stress values. Statistical analysis of inter- and intrareader variability, inter-software and inter-scanner comparison included calculation of absolute and relative error (E), intraclass correlation coefficient (ICC), Bland-Altman analysis, and equivalence testing based on the assumption that inter-software differences needed to be within 80% of the range of intrareader differences.

RESULTS

SW-A and SW-C were the only software programs showing agreement for stroke volume (ICC = 0.96; E = 3 ± 8%), peak flow (ICC: 0.97; E = -1 ± 7%), and area (ICC = 0.81; E = 2 ± 22%). Results from SW-A/D and SW-C/D were equivalent only for area and peak flow. Other software pairs did not yield equivalent results for routinely used clinical parameters. Especially peak maximum velocity yielded poor agreement (ICC ≤ 0.4) between all software packages except SW-A/D that showed good agreement (ICC = 0.80). Inter- and intrareader consistency for clinically used parameters was best for SW-A and SW-D (ICC = 0.56-97) and worst for SW-B (ICC = -0.01-0.71). Of note, inter-scanner differences per individual tended to be smaller than inter-software differences.

CONCLUSIONS

Of all tested software programs, only SW-A and SW-C can be used equivalently for determination of stroke volume, peak flow, and vessel area. Irrespective of the applied software and scanner, high intra- and interreader variability for all parameters have to be taken into account before introducing 4D Flow CMR in clinical routine. Especially in multicenter clinical trials a single image evaluation software should be applied.

摘要

背景

有多种不同的软件程序可用于评估 4D Flow 心血管磁共振(CMR)。程序间结果的良好一致性是该方法被接受的前提。因此,本研究的目的是比较在两个不同供应商的扫描仪上对 8 名健康受试者进行检查,并使用 4 种后处理软件包分析的交叉比较的定量结果。

方法

8 名健康受试者(27±3 岁,3 名女性)分别在两台 3T CMR 系统(Ingenia,飞利浦医疗保健;MAGNETOM Skyra,西门子医疗)上进行标准化的 4D Flow CMR 序列检查。六个手动放置的主动脉轮廓使用 Caas(Pie Medical Imaging,SW-A)、cvi42(Circle Cardiovascular Imaging,SW-B)、GTFlow(GyroTools,SW-C)和 MevisFlow(Fraunhofer Institute MEVIS,SW-D)进行评估,以分析包括心排量、峰值流量、峰值速度和面积在内的 7 个临床常用参数,以及通常用于科学研究的壁面切应力值。使用绝对和相对误差(E)、组内相关系数(ICC)、Bland-Altman 分析和等效性检验(基于软件间差异需要在 80%的读者内差异范围内的假设)对阅读器内、阅读器间、软件间和扫描仪间的差异进行统计分析。

结果

SW-A 和 SW-C 是唯一显示出心排量(ICC=0.96;E=3±8%)、峰值流量(ICC:0.97;E=-1±7%)和面积(ICC=0.81;E=2±22%)一致的软件程序。SW-A/D 和 SW-C/D 的结果仅在面积和峰值流量上等效。其他软件对的常用临床参数结果并不等效。特别是峰值最大速度在所有软件包之间的一致性都很差(ICC≤0.4),除了 SW-A/D 有较好的一致性(ICC=0.80)。在临床常用参数方面,SW-A 和 SW-D 的读者内和读者间一致性最好(ICC=0.56-97),SW-B 最差(ICC=-0.01-0.71)。值得注意的是,个体间的扫描仪差异往往小于软件间的差异。

结论

在所测试的软件程序中,只有 SW-A 和 SW-C 可用于确定心排量、峰值流量和血管面积。在将 4D Flow CMR 引入临床常规之前,应考虑到所有参数的高读者内和读者间变异性,无论应用的软件和扫描仪如何。特别是在多中心临床试验中,应使用单一的图像评估软件。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/db41/10052852/e0f5d41a5aa3/12968_2023_921_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验