Stanford University, Edward L. Ginzton Laboratory, Stanford, California, United States.
SCREEN Holdings Co., Ltd., R&D Department, Kyoto, Japan.
J Biomed Opt. 2019 Apr;24(4):1-6. doi: 10.1117/1.JBO.24.4.046501.
We demonstrate improved optical sectioning in light sheet fluorescence microscopy using tunable structured illumination (SI) frequencies to optimize image quality in scattering specimens. The SI patterns are generated coherently using a one-dimensional spatial light modulator for maximum pattern contrast, and the pattern spatial frequency is adjustable up to half the incoherent cutoff frequency of our detection objective. At this frequency, we demonstrate background reductions of 2 orders of magnitude.