Dispersive viscosurgical devices demonstrate greater efficacy in protecting corneal endothelium in vitro.

作者信息

Yildirim Timur Mert, Auffarth Gerd U, Son Hyeck-Soo, Khoramnia Ramin, Munro Donald John, Merz Patrick R

机构信息

The David J Apple International Laboratory for Ocular Pathology, Heidelberg University Eye Clinic, Heidelberg, Germany.

出版信息

BMJ Open Ophthalmol. 2019 Feb 16;4(1):e000227. doi: 10.1136/bmjophth-2018-000227. eCollection 2019.

Abstract

OBJECTIVE

During phacoemulsification, the corneal endothelium is protected by an ophthalmic viscosurgical device (OVD). In this in vitro study, we assessed six different OVDs for their effectiveness in protecting the corneal endothelium.

METHODS AND ANALYSIS

Phacoemulsification was performed in cadaver eyes of young pigs. Five syringe units of six different OVDs were tested (Healon EndoCoat, Viscoat, Methylvisc, Healon, Healon GV, ProVisc). After surgery, the area of endothelium coated with OVD was determined in relation to the total endothelial surface. Additionally, an endothelial cell count was obtained. As a control, an endothelial cell count was obtained from freshly trephined corneas. Statistical analysis was performed using the Mann-Whitney U test and the Spearman correlation.

RESULTS

The least postoperative endothelial coating and cell count were observed in the cohesive OVDs while the dispersive OVDs showed statistically significant higher values. Healon EndoCoat and Viscoat yielded a coating area of 86 (85-92)% and 85 (85-90)%, respectively. Endothelial cell count was highest in the two dispersive groups with 4065 (3928-4088) cells/mm (Methylvisc) and 4032 (4015-4115) cells/mm (Viscoat). Endothelial coating area and endothelial cell count correlated statistically significantly.

CONCLUSION

Dispersive OVDs from this study showed greater adherence to the endothelial surface than the cohesive ones. Furthermore, postoperative endothelial cell counts of corneas treated with dispersive OVDs were higher than of corneas treated with cohesive OVDs. Our in vitro results suggest that dispersive OVDs protect the corneal endothelium better during phacoemulsification than cohesive OVDs.

摘要

目的

在超声乳化手术中,角膜内皮由眼科粘弹剂(OVD)保护。在这项体外研究中,我们评估了六种不同的OVD对角膜内皮的保护效果。

方法与分析

在幼猪的尸体眼上进行超声乳化手术。测试了六种不同OVD的五个注射器单元(Healon EndoCoat、Viscoat、Methylvisc、Healon、Healon GV、ProVisc)。手术后,确定涂有OVD的内皮面积与总内皮表面积的关系。此外,还进行了内皮细胞计数。作为对照,从新鲜环钻的角膜中获取内皮细胞计数。使用曼-惠特尼U检验和斯皮尔曼相关性进行统计分析。

结果

在粘性OVD中观察到术后内皮覆盖和细胞计数最少,而分散性OVD显示出统计学上显著更高的值。Healon EndoCoat和Viscoat的覆盖面积分别为86(85 - 92)%和85(85 - 90)%。两个分散性组的内皮细胞计数最高,分别为4065(3928 - 4088)个细胞/mm²(Methylvisc)和4032(4015 - 4115)个细胞/mm²(Viscoat)。内皮覆盖面积与内皮细胞计数具有统计学显著相关性。

结论

本研究中的分散性OVD比粘性OVD对内皮表面的粘附性更强。此外,用分散性OVD处理的角膜术后内皮细胞计数高于用粘性OVD处理的角膜。我们的体外研究结果表明,在超声乳化手术中,分散性OVD比粘性OVD能更好地保护角膜内皮。