可疑乳腺肿块的自动与手持式乳腺超声检查的可靠性

Reliability of automated versus handheld breast ultrasound examinations of suspicious breast masses.

作者信息

Yun Gabin, Kim Sun Mi, Yun Bo La, Ahn Hye Shin, Jang Mijung

机构信息

Department of Radiology, Seoul National University Bundang Hospital, Seongnam, Korea.

Department of Radiology, Chung-Ang University Hospital, Chung-Ang University College of Medicine, Seoul, Korea.

出版信息

Ultrasonography. 2019 Jul;38(3):264-271. doi: 10.14366/usg.18055. Epub 2018 Dec 23.

Abstract

PURPOSE

The purpose of this study was to assess the reliability of automated breast ultrasound (ABUS) examinations of suspicious breast masses in comparison to handheld breast ultrasound (HHUS) with regard to Breast Imaging Reporting and Data System (BI-RADS) category assessment, and to investigate the factors affecting discrepancies in categorization.

METHODS

A total of 135 masses that were assessed as BI-RADS categories 4 and 5 on ABUS that underwent ultrasound (US)-guided core needle biopsy from May 2017 to December 2017 were included in this study. The BI-RADS categories were re-assessed using HHUS. Agreement of the BI-RADS categories was evaluated using kappa statistics, and the positive predictive value of each examination was calculated. Logistic regression analysis was performed to identify the mammography and US findings associated with discrepancies in the BI-RADS categorization.

RESULTS

The overall agreement between ABUS and HHUS in all cases was good (79.3%, kappa=0.61, P<0.001). Logistic regression analysis revealed that accompanying suspicious microcalcifications on mammography (odds ratio [OR], 4.63; 95% confidence interval [CI], 1.83 to 11.71; P=0.001) and an irregular shape on US (OR, 5.59; 95% CI, 1.43 to 21.83; P=0.013) were associated with discrepancies in the BI-RADS categorization.

CONCLUSION

The agreement between ABUS and HHUS examinations in the BI-RADS categorization of suspicious breast masses was good. The presence of suspicious microcalcifications on mammography and an irregular shape on US were factors associated with ABUS yielding a lower level of suspicion than HHUS in terms of the BI-RADS category assessment.

摘要

目的

本研究旨在评估自动乳腺超声(ABUS)检查可疑乳腺肿块与手持乳腺超声(HHUS)相比,在乳腺影像报告和数据系统(BI-RADS)分类评估方面的可靠性,并调查影响分类差异的因素。

方法

本研究纳入了2017年5月至2017年12月期间在ABUS上被评估为BI-RADS 4类和5类且接受了超声(US)引导下粗针活检的135个肿块。使用HHUS对BI-RADS类别进行重新评估。使用kappa统计量评估BI-RADS类别的一致性,并计算每次检查的阳性预测值。进行逻辑回归分析以确定与BI-RADS分类差异相关的乳腺X线摄影和超声检查结果。

结果

所有病例中ABUS和HHUS之间的总体一致性良好(79.3%,kappa = 0.61,P < 0.001)。逻辑回归分析显示,乳腺X线摄影上伴有可疑微钙化(比值比[OR],4.63;95%置信区间[CI],1.83至11.71;P = 0.001)以及超声检查时形状不规则(OR,5.59;95% CI,1.43至21.83;P = 0.013)与BI-RADS分类差异相关。

结论

ABUS和HHUS检查在可疑乳腺肿块的BI-RADS分类方面一致性良好。乳腺X线摄影上存在可疑微钙化以及超声检查时形状不规则是在BI-RADS类别评估中ABUS比HHUS产生较低可疑程度的相关因素。