Case Center for Imaging Research, Case Western Reserve University, Cleveland, OH, USA.
Department of Radiology, Case Western Reserve University, Cleveland, OH, USA.
Med Phys. 2019 Aug;46(8):3520-3531. doi: 10.1002/mp.13574. Epub 2019 Jun 12.
Accurate photon attenuation assessment from MR data remains an unmet challenge in the thorax due to tissue heterogeneity and the difficulty of MR lung imaging. As thoracic tissues encompass the whole physiologic range of photon absorption, large errors can occur when using, for example, a uniform, water-equivalent or a soft-tissue-only approximation. The purpose of this study was to introduce a method for voxel-wise thoracic synthetic CT (sCT) generation from MR data attenuation correction (AC) for PET/MR or for MR-only radiation treatment planning (RTP).
Acquisition: A radial stack-of-stars combining ultra-short-echo time (UTE) and modified Dixon (mDixon) sequence was optimized for thoracic imaging. The UTE-mDixon pulse sequence collects MR signals at three TE times denoted as UTE, Echo1, and Echo2. Three-point mDixon processing was used to reconstruct water and fat images. Bias field correction was applied in order to avoid artifacts caused by inhomogeneity of the MR magnetic field.
Water fraction and R2* maps were estimated using the UTE-mDixon data to produce a total of seven MR features, that is UTE, Echo1, Echo2, Dixon water, Dixon fat, Water fraction, and R2*. A feature selection process was performed to determine the optimal feature combination for the proposed automatic, 6-tissue classification for sCT generation. Fuzzy c-means was used for the automatic classification which was followed by voxel-wise attenuation coefficient assignment as a weighted sum of those of the component tissues. Performance evaluation: MR data collected using the proposed pulse sequence were compared to those using a traditional two-point Dixon approach. Image quality measures, including image resolution and uniformity, were evaluated using an MR ACR phantom. Data collected from 25 normal volunteers were used to evaluate the accuracy of the proposed method compared to the template-based approach. Notably, the template approach is applicable here, that is normal volunteers, but may not be robust enough for patients with pathologies.
The free breathing UTE-mDixon pulse sequence yielded images with quality comparable to those using the traditional breath holding mDixon sequence. Furthermore, by capturing the signal before T2* decay, the UTE-mDixon image provided lung and bone information which the mDixon image did not. The combination of Dixon water, Dixon fat, and the Water fraction was the most robust for tissue clustering and supported the classification of six tissues, that is, air, lung, fat, soft tissue, low-density bone, and dense bone, used to generate the sCT. The thoracic sCT had a mean absolute difference from the template-based (reference) CT of less than 50 HU and which was better agreement with the reference CT than the results produced using the traditional Dixon-based data.
MR thoracic acquisition and analyses have been established to automatically provide six distinguishable tissue types to generate sCT for MR-based AC of PET/MR and for MR-only RTP.
由于组织异质性和肺部磁共振成像的困难,准确评估胸部的光子衰减仍然是一个未满足的挑战。由于胸部组织涵盖了光子吸收的整个生理范围,因此当使用例如均匀的、水等效的或仅软组织的近似值时,会产生很大的误差。本研究的目的是介绍一种从 MR 数据衰减校正 (AC) 生成用于 PET/MR 或用于仅 MR 放射治疗计划 (RTP) 的胸部合成 CT (sCT) 的体素级方法。
采集:采用径向星形堆叠技术,结合超短回波时间 (UTE) 和改良 Dixon (mDixon) 序列,对胸部进行了优化成像。UTE-mDixon 脉冲序列在三个 TE 时间收集 MR 信号,分别标记为 UTE、Echo1 和 Echo2。使用三点 Dixon 处理来重建水和脂肪图像。应用偏置场校正以避免由 MR 磁场不均匀引起的伪影。
使用 UTE-mDixon 数据估计水分数和 R2图,以产生总共七种 MR 特征,即 UTE、Echo1、Echo2、Dixon 水、Dixon 脂肪、水分数和 R2。进行特征选择过程以确定用于自动、6 组织分类的最佳特征组合,以生成 sCT。使用模糊 c-均值进行自动分类,然后将衰减系数分配给各组织的加权和。性能评估:比较了使用提出的脉冲序列和传统两点 Dixon 方法采集的 MR 数据。使用 MR ACR 体模评估图像质量指标,包括图像分辨率和均匀性。使用 25 名正常志愿者的数据来评估与基于模板的方法相比,所提出方法的准确性。值得注意的是,该模板方法适用于正常志愿者,但对于患有病理的患者可能不够稳健。
自由呼吸的 UTE-mDixon 脉冲序列产生的图像质量与使用传统呼吸暂停 Dixon 序列相当。此外,通过捕获 T2*衰减之前的信号,UTE-mDixon 图像提供了肺和骨骼信息,而 Dixon 图像没有。Dixon 水、Dixon 脂肪和水分数的组合最适合组织聚类,并支持生成 sCT 的六种组织的分类,即空气、肺、脂肪、软组织、低密度骨和高密度骨。与基于模板的(参考)CT 相比,胸部 sCT 的平均绝对差异小于 50HU,与参考 CT 的一致性优于使用传统 Dixon 数据生成的结果。
已经建立了用于 MR 胸部采集和分析的方法,以自动提供六种可区分的组织类型,从而为基于 MR 的 PET/MR 和仅用于 MR 的 RTP 的 AC 生成 sCT。