Ustundağ Yasemin, Huysal Kağan, Ozgunay Şeyda E, Turkoğlu Ali R
Department of Clinical Biochemistry, University of Health Sciences, Bursa YuksekIhtisas Training and Research Hospital, Bursa, Turkey.
Department of Anestesiology and Reanimation, University of Health Sciences, Bursa YuksekIhtisas Training and Research Hospital, Bursa, Turkey.
Indian J Crit Care Med. 2019 Jan;23(1):35-42. doi: 10.5005/jp-journals-10071-23110.
The major extracellular electrolytes, sodium, and potassium are often requested together and form a large percentage of the requested tests in routine clinical chemistry laboratories. Two types of devices that use direct and indirect ion-selective electrode (ISE) methods are used in hospitals for electrolyte measurements: blood gas analyzers (BGA), which use direct ISE technology, and the indirect ISE method, which is often used in a central-laboratory autoanalyzer (AA). We aimed to summarize the current scientific knowledge based on whether the electrolyte test results, using Na and K test results obtained with BGA and an AA, can be used interchangeably. We searched Medline (PubMed), Google Scholar, and Web of Science up to 31st March 2018. In addition, references of the included studies were also examined. Fourteen studies with a risk of bias were included in the analysis. Limits of agreement differences were variable among BGA and AA sodium and potassium test results in clinical practice. The results of both BGA and AA measures should not be used interchangeably under the assumption that they are equivalent to each other.
Ustundağ Y, Huysal K, Ozgunay ŞE, Turkoğlu AR. Interchangeability of Sodium and Potassium Result Values of Arterial Blood Gas with Laboratory Analyzer: Narrative Review. Indian Journal of Critical Care Medicine, January 2019;23(1):35-42.
主要的细胞外电解质钠和钾经常被一起检测,并且在常规临床化学实验室所检测的项目中占很大比例。医院中用于电解质测量的设备有两种,分别采用直接和间接离子选择电极(ISE)方法:使用直接ISE技术的血气分析仪(BGA),以及常用于中心实验室自动分析仪(AA)的间接ISE方法。我们旨在根据使用BGA和AA获得的钠和钾检测结果所得到的电解质检测结果是否可以互换使用,来总结当前的科学知识。我们检索了截至2018年3月31日的Medline(PubMed)、谷歌学术和科学网。此外,还查阅了纳入研究的参考文献。分析纳入了14项存在偏倚风险的研究。在临床实践中,BGA和AA的钠和钾检测结果之间的一致性差异限度各不相同。在假设BGA和AA的测量结果彼此等效的情况下,两者的结果不应互换使用。
Ustundağ Y, Huysal K, Ozgunay ŞE, Turkoğlu AR. 动脉血气与实验室分析仪钠和钾结果值的互换性:叙述性综述。《印度重症医学杂志》,2019年1月;23(1):35 - 42。