Suppr超能文献

不同往复运动系统在不同工作长度下对弯曲根管内碎屑推出效果的研究。

Extrusion of Debris from Curved Root Canals Instrumented up to Different Working Lengths Using Different Reciprocating Systems.

机构信息

Department of Endodontics, São Leopoldo Mandic Dental Research Center, Campinas, São Paulo, Brazil.

Department of Endodontics, São Leopoldo Mandic Dental Research Center, Campinas, São Paulo, Brazil.

出版信息

J Endod. 2019 Jul;45(7):930-934. doi: 10.1016/j.joen.2019.03.002. Epub 2019 May 17.

Abstract

INTRODUCTION

The aim of this study was to evaluate the extrusion of debris associated with the use of the Reciproc (VDW GmbH, Munich, Germany) and WaveOne Gold (Dentsply Maillefer, Ballaigues, Switzerland) reciprocating systems in curved canals instrumented up to different working lengths (WLs).

METHODS

Sixty mesial roots of extracted mandibular first molars with mesiobuccal canals having curvatures between 10° and 20° and independent foramina were randomly distributed into 4 experimental groups (n = 15) according to WL and the instrumentation system used: RC0, Reciproc and WL at the apical foramen; RC1, Reciproc and WL at 1 mm short of the apical foramen; WOG0, WaveOne Gold and WL at the apical foramen; and WOG1, WaveOne Gold and WL at 1 mm short of the apical foramen. The canals were irrigated with double-distilled water throughout the preparation. Each root was fit into an Eppendorf tube to collect the debris extruded through the foramen during instrumentation. After instrumentation, the roots were removed from the tubes, and the weight of the debris was obtained by subtracting the initial weight of the tubes from their final weight. The data were submitted to the analysis of variance parametric test complemented by the Tukey test.

RESULTS

No statistically significant difference was observed between the 4 experimental groups regarding the amount of debris extruded (P > .05).

CONCLUSIONS

The 2 reciprocating systems tested were associated with similar amounts of apically extruded debris, regardless of the working length used in the instrumentation.

摘要

简介

本研究旨在评估在弯曲根管中使用 Reciproc(VDW GmbH,慕尼黑,德国)和 WaveOne Gold(Dentsply Maillefer,Ballaigues,瑞士)往复式系统到达不同工作长度(WL)时与器械相关的碎屑挤出量。

方法

将 60 颗具有 10°至 20°弯曲度和独立根尖孔的下颌第一磨牙近中根随机分为 4 个实验组(n = 15),根据 WL 和使用的器械系统:RC0,Reciproc 和 WL 在根尖孔处;RC1,Reciproc 和 WL 在根尖孔 1mm 处;WOG0,WaveOne Gold 和 WL 在根尖孔处;和 WOG1,WaveOne Gold 和 WL 在根尖孔 1mm 处。整个预备过程中用双蒸水冲洗根管。每个根管都被放置在一个 Eppendorf 管中,以收集在器械预备过程中从根尖孔挤出的碎屑。器械预备完成后,将根管从管中取出,通过用最终重量减去初始重量来获得碎屑的重量。将数据提交方差分析参数检验,并通过 Tukey 检验进行补充。

结果

4 个实验组之间在挤出的碎屑量方面没有统计学差异(P >.05)。

结论

无论使用的工作长度如何,测试的两种往复式系统与相似数量的根尖挤出碎屑相关。