Scheimpflug Camera and Swept-Source Optical Coherence Tomography in Pachymetry Evaluation of Diabetic Patients.

作者信息

Krysik Katarzyna, Dobrowolski Dariusz, Stanienda-Sokół Karolina, Wylegala Edward A, Lyssek-Boron Anita

机构信息

Department of Ophthalmology with Pediatric Unit, St. Barbara Hospital, Trauma Center, Medykow Square 1, 41-200 Sosnowiec, Poland.

Chair and Clinical Department of Ophthalmology, School of Medicine with the Division of Dentistry in Zabrze, Medical University of Silesia in Katowice, Panewnicka 65 St., 40-760 Katowice, Poland.

出版信息

J Ophthalmol. 2019 Apr 15;2019:4532657. doi: 10.1155/2019/4532657. eCollection 2019.

Abstract

AIM

The comparative analysis of the central and peripheral corneal thicknesses using two different imaging systems: Scheimpflug camera and swept-source OCT was performed to investigate the differences in corneal thickness analysis in diabetic patients.

MATERIALS AND METHODS

The study group consisted of the 147 eyes of 107 diabetic patients who were examined and compared with 138 eyes of 89 nondiabetic cataract patients. The inclusion criteria for the study group was diabetes mellitus type II identified no less than 10 years ago, with NPDR not requiring prior laser treatment. The control group was recruited from nondiabetic patients. Measurements were obtained on the Pentacam Scheimpflug imaging system and Casia swept-source OCT. All study parameters from anterior chamber images were processed for five different zones, the central zone and four peripherals-superior, inferior, nasal, and temporal. A fit zone diameter of 4 mm was applied for both instruments.

RESULTS

The Pentacam system overestimated corneal measurements in the DM group when compared with the Casia OCT in superior corneal zone (=0.04), inferior corneal zone (=0.02), nasal corneal zone ( < 0.001), and temporal corneal zone (=0.01). In the control group, there were also statistically significant differences between the Pentacam and Casia OCT measured values in inferior corneal zone (=0.001), nasal corneal zone (=0.04), and temporal corneal zone ( < 0.001).

CONCLUSION

Scheimpflug camera pachymetry measurements showed statistically higher CCT values when compared with swept-source OCT measurements.

摘要

目的

使用两种不同的成像系统——眼前节分析系统和扫频光学相干断层扫描(OCT)对中央和周边角膜厚度进行比较分析,以研究糖尿病患者角膜厚度分析的差异。

材料与方法

研究组由107例糖尿病患者的147只眼组成,对其进行检查并与89例非糖尿病白内障患者的138只眼进行比较。研究组的纳入标准为10年前确诊的II型糖尿病,非增殖性糖尿病视网膜病变(NPDR)且无需先前的激光治疗。对照组从非糖尿病患者中招募。使用Pentacam眼前节分析成像系统和Casia扫频OCT进行测量。对前房图像的所有研究参数在五个不同区域进行处理,即中央区和四个周边区域——上方、下方、鼻侧和颞侧。两种仪器均采用4mm的拟合区直径。

结果

与Casia OCT相比,Pentacam系统在糖尿病组上方角膜区(=0.04)、下方角膜区(=0.02)、鼻侧角膜区(<0.001)和颞侧角膜区(=0.01)高估了角膜测量值。在对照组中,Pentacam和Casia OCT测量值在下方角膜区(=0.001)、鼻侧角膜区(=0.04)和颞侧角膜区(<0.001)也存在统计学显著差异。

结论

与扫频OCT测量相比,眼前节分析系统测量的中央角膜厚度(CCT)值在统计学上更高。