Electron backscatter corrections for parallel-plate chambers.

作者信息

Hunt M A, Kutcher G J, Buffa A

机构信息

Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, New York 10021.

出版信息

Med Phys. 1988 Jan-Feb;15(1):96-103. doi: 10.1118/1.596165.

Abstract

The wall of an ionization chamber is commonly assumed to have a negligible effect on chamber response in electron beams. For cylindrical chambers with thin walls this assumption is valid. However, parallel-plate chambers commonly possess large mechanical supports which may affect chamber response in a manner not accounted for in current dosimetry protocols. This is due to changes with energy in the relative backscattered electron fluence between chamber support and phantom materials. To investigate this effect, electron backscatter from low atomic number materials has been measured with electrons from 6 to 20 MeV. The effect of the diameter and thickness of the backscattering material has also been studied. Based on these data, Lucite and polystyrene chambers in water phantoms are expected to underrespond by 1% and 2% at 6 MeV. The expected underresponse decreases to 0.8% and 0.4% for polystyrene and Lucite at 12 MeV and is insignificant above 16 MeV. Two commercially available parallel-plate chambers were compared with a cylindrical chamber in electron beams from 6 to 20 MeV. Using the 20-MeV intercomparison, the expected chamber responses at the lower energies were calculated and compared with measurements. Both parallel-plate chambers underresponded by approximately 1% at 6 MeV and 0.5% at 9 MeV which is qualitatively consistent with the electron backscatter data. Recommendations for minimizing electron backscatter effects through chamber design are discussed.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验