Quantification of porcine myocardial perfusion with modified dual bolus MRI - a prospective study with a PET reference.

机构信息

Diagnostic Imaging Center, Kuopio University Hospital, PO Box 100, 70029, Kuopio, KYS, Finland.

Department of Applied Physics, University of Eastern Finland, Kuopio, Finland.

出版信息

BMC Med Imaging. 2019 Jul 26;19(1):58. doi: 10.1186/s12880-019-0359-8.

Abstract

BACKGROUND

The reliable quantification of myocardial blood flow (MBF) with MRI, necessitates the correction of errors in arterial input function (AIF) caused by the T1 saturation effect. The aim of this study was to compare MBF determined by a traditional dual bolus method against a modified dual bolus approach and to evaluate both methods against PET in a porcine model of myocardial ischemia.

METHODS

Local myocardial ischemia was induced in five pigs, which were subsequently examined with contrast enhanced MRI (gadoteric acid) and PET (O-15 water). In the determination of MBF, the initial high concentration AIF was corrected using the ratio of low and high contrast AIF areas, normalized according to the corresponding heart rates. MBF was determined from the MRI, during stress and at rest, using the dual bolus and the modified dual bolus methods in 24 segments of the myocardium (total of 240 segments, five pigs in stress and rest). Due to image artifacts and technical problems 53% of the segments had to be rejected from further analyses. These two estimates were later compared against respective rest and stress PET-based MBF measurements.

RESULTS

Values of MBF were determined for 112/240 regions. Correlations for MBF between the modified dual bolus method and PET was r = 0.84, and between the traditional dual bolus method and PET r = 0.79. The intraclass correlation was very good (ICC = 0.85) between the modified dual bolus method and PET, but poor between the traditional dual bolus method and PET (ICC = 0.07).

CONCLUSIONS

The modified dual bolus method showed a better agreement with PET than the traditional dual bolus method. The modified dual bolus method was found to be more reliable than the traditional dual bolus method, especially when there was variation in the heart rate. However, the difference between the MBF values estimated with either of the two MRI-based dual-bolus methods and those estimated with the gold-standard PET method were statistically significant.

摘要

背景

利用 MRI 可靠地量化心肌血流(MBF),需要校正动脉输入函数(AIF)中的 T1 饱和效应引起的误差。本研究旨在比较传统双脉冲法和改良双脉冲法测定的 MBF,并在心肌缺血的猪模型中分别用这两种方法与 PET 进行比较。

方法

在 5 头猪中诱发局部心肌缺血,随后用对比增强 MRI(钆喷替酸葡甲胺)和 PET(O-15 水)进行检查。在 MBF 的测定中,使用高低对比度 AIF 区域比值校正初始高浓度 AIF,根据相应的心率进行归一化。在应激和休息时,使用双脉冲和改良双脉冲法,通过 24 个心肌节段(5 头猪在应激和休息时的总共 240 个节段)测定 MRI 的 MBF。由于图像伪影和技术问题,53%的节段必须从进一步的分析中排除。这两个估计值后来与各自的静息和应激 PET 基础 MBF 测量值进行了比较。

结果

确定了 240 个节段中的 112 个区域的 MBF 值。改良双脉冲法与 PET 之间的 MBF 相关性为 r=0.84,传统双脉冲法与 PET 之间的相关性为 r=0.79。改良双脉冲法与 PET 的组内相关系数非常好(ICC=0.85),但传统双脉冲法与 PET 的组内相关系数较差(ICC=0.07)。

结论

改良双脉冲法与 PET 的一致性优于传统双脉冲法。改良双脉冲法比传统双脉冲法更可靠,尤其是在心率变化时。然而,用两种 MRI 双脉冲法中的任何一种方法估计的 MBF 值与用金标准 PET 法估计的 MBF 值之间存在统计学差异。