Suppr超能文献

角膜可视化Scheimpflug技术测量的生物力学校正眼压与传统非接触眼压测量的对比分析

Comparative analysis of biomechanically corrected intraocular pressure with corneal visualization Scheimpflug technology versus conventional noncontact intraocular pressure.

作者信息

Ma Jiaonan, Wang Yan, Hao Weiting, Jhanji Vishal

机构信息

Clinical College of Ophthalmology, Tianjin Medical University, No. 4. Gansu Road, He-ping District, Tianjin, 300020, China.

Tianjin Eye Hospital, Tianjin Eye Institute, Tianjin Key Laboratory of Ophthalmology and Visual Science, Nankai University Affiliated Eye Hospital, Tianjin, China.

出版信息

Int Ophthalmol. 2020 Jan;40(1):117-124. doi: 10.1007/s10792-019-01159-9. Epub 2019 Aug 20.

Abstract

PURPOSE

To compare the difference between biomechanically corrected intraocular pressure (bIOP) and noncontact IOP measurement (IOP) and to investigate the effect of corneal biomechanical properties on IOP.

METHODS

IOP was evaluated in 1046 myopic eyes (544 subjects) using a conventional noncontact tonometer and a novel corneal visualization Scheimpflug technology (Corvis ST). Corneal biomechanical parameters were measured using the Corvis ST.

RESULTS

The mean IOP and bIOP values were significantly different (15.59 ± 2.56 mmHg and 15.89 ± 1.75 mmHg, respectively; P < 0.001). The bIOP showed a less correlation with central corneal thickness (CCT), compared with IOP (P < 0.01). The IOP was lower than the bIOP when the thickness of cornea was ≤ 550 μm but higher than bIOP when it was ≥ 550 μm (P < 0.01). A strong association was found between IOP and deflection amplitude and deflection area at the highest concavity (HC DefA and HC DefArea), stiff parameter, maximum deformation amplitude (DAmax), and maximum deflection amplitude (DefAmax), as well as for bIOP (r > 0.500, P < 0.001). The bIOP could be calculated based on IOP according to different values of CCT (P < 0.01).

CONCLUSIONS

The bIOP was less affected by CCT as compared to IOP. IOP may be underestimated when the cornea is thinner and overestimated when the cornea is thicker because of the difference in corneal biomechanics.

摘要

目的

比较生物力学校正眼压(bIOP)与非接触眼压测量值(IOP)之间的差异,并研究角膜生物力学特性对眼压的影响。

方法

使用传统非接触眼压计和新型角膜可视化Scheimpflug技术(Corvis ST)对1046只近视眼(544名受试者)进行眼压评估。使用Corvis ST测量角膜生物力学参数。

结果

平均眼压和bIOP值存在显著差异(分别为15.59±2.56 mmHg和15.89±1.75 mmHg;P<0.001)。与眼压相比,bIOP与中央角膜厚度(CCT)的相关性较小(P<0.01)。当角膜厚度≤550μm时,眼压低于bIOP;而当角膜厚度≥550μm时,眼压高于bIOP(P<0.01)。眼压与最高凹度处的偏转幅度和偏转面积(HC DefA和HC DefArea)、硬度参数、最大变形幅度(DAmax)和最大偏转幅度(DefAmax)之间存在强关联,bIOP也如此(r>0.500,P<0.001)。根据不同的CCT值,可基于眼压计算bIOP(P<0.01)。

结论

与眼压相比,bIOP受CCT的影响较小。由于角膜生物力学的差异,角膜较薄时眼压可能被低估,而角膜较厚时眼压可能被高估。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验