To evaluate the Spot Vision Screener in detecting targeted vision disorders compared to cycloplegic retinoscopy in children <3 years of age.
Children, ages 6 months to 36 months underwent vision screening using the Spot Vision Screener. Results were compared to results of comprehensive eye examinations. Validity of the Spot was evaluated by calculating the area under the curve (AUC); the receiver operating characteristics (ROC) were used to determine optimal sensitivity and specificity for detection of targeted vision disorders.
A total of 249 children were included. The AUC for detecting targeted vision disorders as defined by the study specific criteria using the Spot was 0.790. Compared to cycloplegic retinoscopy, the Spot underestimated hyperopia by 1.02 D (95% CI, 0.86-1.17 D). For hyperopia ≥4.5 D spherical equivalent (n = 10), the mean difference between the Spot and cycloplegic retinoscopy was 3.46 D (95% CI, 1.95-4.98 D). In contrast, the Spot overestimated astigmatism compared to cycloplegic retinoscopy (-1.00 D vs -0.48 D; P < 0.001) by -0.52 D (95% CI, 0.43-0.62 D).
The Spot Vision Screener showed good overall validity in detecting targeted vision disorders. It was within 0.5 D and 1 D of cycloplegic retinoscopy with regard to low hyperopia and astigmatism. Higher hyperopic spherical equivalent refractive errors showed larger differences in mean values between the Spot and cycloplegic retinoscopy.
评估 Spot 视力筛查仪在检测 <3 岁儿童目标视力障碍方面的表现,与睫状肌麻痹验光相比。
6 个月至 36 个月大的儿童接受 Spot 视力筛查仪进行视力筛查。结果与全面眼科检查结果进行比较。通过计算曲线下面积(AUC)来评估 Spot 的有效性;使用接收者操作特征(ROC)确定检测目标视力障碍的最佳敏感性和特异性。
共纳入 249 名儿童。使用 Spot 按照研究特定标准检测目标视力障碍的 AUC 为 0.790。与睫状肌麻痹验光相比,Spot 低估远视 1.02 D(95%置信区间,0.86-1.17 D)。对于等效球镜远视≥4.5 D(n=10),Spot 与睫状肌麻痹验光之间的平均差异为 3.46 D(95%置信区间,1.95-4.98 D)。相比之下,Spot 高估了散光,与睫状肌麻痹验光相比,差值为-1.00 D(95%置信区间,-0.48 D 至-0.48 D)。
Spot 视力筛查仪在检测目标视力障碍方面具有良好的总体有效性。在低度远视和散光方面,它与睫状肌麻痹验光的差值在 0.5 D 和 1 D 以内。较高的远视等效球镜屈光不正显示 Spot 与睫状肌麻痹验光之间的平均值差异较大。