This study aimed to explore the accuracy and precision of three-dimensional optical (3DO) whole-body scanning for automated anthropometry and estimating total and regional body composition.
Healthy children and adolescents (n = 181, ages 5-17 years) were recruited for the Shape Up! Kids study. Each participant underwent whole-body dual-energy x-ray absorptiometry and 3DO scans; multisite conventional tape measurements served as the anthropometric criterion measure. 3DO body shape was described using automated body circumference, length, and volume measures. 3DO estimates were compared with criterion measures using simple linear regression by the stepwise selection method.
Of the 181 participants, 112 were used for the training set, 49 were used for the test set, and 20 were excluded for technical reasons. 3DO body composition estimates were strongly associated with dual-energy x-ray absorptiometry measures for percent body fat, fat mass, and fat-free mass (R : 0.83, 0.96, and 0.98, respectively). 3DO provided reliable measurements of fat mass (coefficient of variation, 3.30; root mean square error [RMSE], 0.53), fat-free mass (coefficient of variation, 1.34; RMSE, 0.53 kg), and percent body fat (RMSE = 1.2%).
3DO surface scanning provides accurate and precise anthropometric and body composition estimates in children and adolescents with high precision. 3DO is a safe, accessible, and practical method for evaluating body shape and composition in research and clinical settings.
本研究旨在探讨三维光学(3DO)全身扫描在自动人体测量和估计总体及局部身体成分方面的准确性和精密度。
健康的儿童和青少年(n=181,年龄 5-17 岁)被招募参加 Shape Up!Kids 研究。每位参与者均接受全身双能 X 射线吸收法和 3DO 扫描;多部位常规胶带测量作为人体测量学的标准测量方法。3DO 身体形状采用自动身体周长、长度和体积测量来描述。采用逐步选择法的简单线性回归比较 3DO 估计值与标准测量值。
在 181 名参与者中,112 名用于训练集,49 名用于测试集,20 名因技术原因被排除在外。3DO 身体成分估计值与双能 X 射线吸收法测量的体脂百分比、脂肪量和去脂体重高度相关(R:分别为 0.83、0.96 和 0.98)。3DO 能够可靠地测量脂肪量(变异系数,3.30;均方根误差[RMSE],0.53)、去脂体重(变异系数,1.34;RMSE,0.53kg)和体脂百分比(RMSE=1.2%)。
3DO 表面扫描可为儿童和青少年提供准确且精确的人体测量和身体成分估计值,具有高精度。3DO 是一种安全、可及且实用的方法,可用于研究和临床环境中评估体型和身体成分。