Here we report on the comprehensive quality control of a 4.04 GBq (109 mCi) generator supplied by itG (Munich, Germany), and used for routine production of [Ga]Ga-PSMA-11 for clinical imaging. The performance of the 4.04 GBq itG Ge/Ga generator was studied for a year and parameters including elution yield, elution profile, radioactive and stable contaminants were collected. The production yields of a series of 175 [Ga]Ga-PSMA-11 clinical batches are also reported herein.
This first-of-its-kind GMP grade Ge/Ga generator from itG with a nominal activity of 4.04 GBq (109 mCi) showed a stable Ga elution profile with elution efficiency averaging 58.3 ± 3.7%. Ge contaminant in the eluent slightly increased over time but remained 100x lower than those reported for comparable 1.85 GBq (50 mCi) itG generators. Metal impurities were found in concentrations lower than 100 ng/ml (ppb) throughout the study. [Ga]Ga-PSMA-11 was obtained in 89 ± 4% radiochemical yields and > 99% radiochemical and chemical purities.
4.04 GBq (109 mCi) itG Ge/Ga generator is suitable for routinely produced Ga tracers used in the clinic. Up to 30% higher amount of final drug product was obtained as compared to the 1.85 GBq (50 mCi) itG generator, and as a result larger number of studies could be performed, while reducing the synthetic burden.
在此,我们报告了由德国慕尼黑的itG公司提供的一台4.04 GBq(109 mCi)发生器的全面质量控制情况,该发生器用于[镓]Ga-PSMA-11的常规生产以用于临床成像。对4.04 GBq的itG锗/镓发生器的性能进行了为期一年的研究,并收集了包括洗脱产率、洗脱曲线、放射性和稳定污染物在内的参数。本文还报告了一系列175个[镓]Ga-PSMA-11临床批次的生产产率。
这台itG公司生产的首台GMP级锗/镓发生器,标称活度为4.04 GBq(109 mCi),其镓洗脱曲线稳定,洗脱效率平均为58.3±3.7%。洗脱液中的锗污染物随时间略有增加,但仍比报道的同类1.85 GBq(50 mCi)itG发生器低100倍。在整个研究过程中,发现金属杂质的浓度低于100 ng/ml(ppb)。[镓]Ga-PSMA-11的放射化学产率为89±4%,放射化学和化学纯度均>99%。