Development and international multicentre pilot testing of a postal dosimetry audit methodology for high dose rate brachytherapy.

作者信息

Dimitriadis Alexis, Becker Anna, Chelminski Krzysztof, Kazantsev Pavel, Titovich Egor, Azangwe Godfrey, de la Fuente Rosales Liset, Kellogg Benjamin, Carrara Mauro, Swamidas Jamema

机构信息

International Atomic Energy Agency, Department of Nuclear Sciences and Applications, Division of Human Health, Vienna, Austria.

出版信息

Phys Imaging Radiat Oncol. 2024 Nov 2;32:100665. doi: 10.1016/j.phro.2024.100665. eCollection 2024 Oct.

Abstract

BACKGROUND AND PURPOSE

Dosimetry audits are essential for reducing errors in brachytherapy. A postal dosimetry audit methodology was developed and tested in an international multicentre pilot, to assess the accuracy of the Reference Air Kerma Rate of Ir and Co brachytherapy sources.

MATERIALS AND METHODS

A compact phantom made of polymethyl methacrylate was developed to accommodate two catheters, a radiophotoluminescence dosimeter (RPLD) for dose measurements and a Gafchromic (RTQA2) film strip for source position verification. Deviations of the audit setup from TG-43 conditions were quantified experimentally and compared to previous Monte Carlo (MC) simulations. A measurement uncertainty budget was estimated for the RPLD analysis. The methodology was tested in an international pilot study consisting of 59 dosimeter sets among 48 centres from 11 countries.

RESULTS

The experimental correction factors showed good agreement with previous MC simulations, and the total correction factor accounting for non-water equivalence, lack of scatter and beam quality was found to be 1.029 ± 0.009 for Ir and 1.059 ± 0.007 for Co sources, to be employed in audit measurement. The total uncertainty budget was estimated to be 2.24 % (k = 1). In the multicentre study, the ratio between measured and reported user dose ranged from 0.968 to 1.049, with all irradiated dosimeter sets within ± 5 %, and 54 out of 59 within ± 3 %.

CONCLUSIONS

The methodology was tested in an international multicentre pilot study and has shown good performance validating the uncertainty budget.

摘要

背景与目的

剂量学审核对于减少近距离放射治疗中的误差至关重要。一种邮政剂量学审核方法在一项国际多中心试点中得以开发和测试,以评估铱源和钴源近距离放射治疗参考空气比释动能率的准确性。

材料与方法

研制了一种由聚甲基丙烯酸甲酯制成的紧凑型模体,用于容纳两根导管、一个用于剂量测量的放射光致发光剂量计(RPLD)和一个用于源位置验证的辐射变色(RTQA2)胶片条。通过实验对审核设置与TG - 43条件的偏差进行了量化,并与先前的蒙特卡罗(MC)模拟结果进行了比较。估算了RPLD分析的测量不确定度预算。该方法在一项国际试点研究中进行了测试,该研究由来自11个国家48个中心的59套剂量计组成。

结果

实验校正因子与先前的MC模拟结果显示出良好的一致性,在审核测量中,考虑非水等效性、散射不足和射束质量的总校正因子对于铱源为1.029±0.009,对于钴源为1.059±0.007。总不确定度预算估计为2.24%(k = 1)。在多中心研究中,测量剂量与报告的用户剂量之比在0.968至1.049之间,所有受辐照的剂量计组偏差均在±5%以内,59组中有54组在±3%以内。

结论

该方法在一项国际多中心试点研究中进行了测试,并表现出良好的性能,验证了不确定度预算。