通过两种不同的共振频率分析装置测量皮质骨和种植体设计对牙种植体初期稳定性的影响:一项研究。

Influence of cortical bone and implant design in the primary stability of dental implants measured by two different devices of resonance frequency analysis: An study.

作者信息

Chávarri-Prado David, Brizuela-Velasco Aritza, Diéguez-Pereira Markel, Pérez-Pevida Esteban, Jiménez-Garrudo Antonio, Viteri-Agustín Iratxe, Estrada-Martínez Alejandro, Montalbán-Vadillo Oier

机构信息

Department of Surgery and Medical-Surgical Specialties, Faculty of Medicine and Dentistry, University of Oviedo, Oviedo, Spain.

Department of Surgery, Faculty of Medicine, University of Salamanca, Salamanca, Spain.

出版信息

J Clin Exp Dent. 2020 Mar 1;12(3):e242-e248. doi: 10.4317/jced.56014. eCollection 2020 Mar.

Abstract

BACKGROUND

This study aimed to evaluate the effect of the implant design and the presence of cortical bone in the primary stability, as well as analyze the differences between the stability measurements obtained by two different resonance frequency analysis (RFA) devices.

MATERIAL AND METHODS

A total of 80 Klockner implants of two different models [40 Essential Cone implants (group A) and 40 Vega implants (group B)] were used. The implants were placed in two polyurethane blocks that simulated the mechanical properties of the maxillary bone. One block featured a layer of cortical bone that was absent from the other block. The primary stability of all implants was measured by insertion torque and RFA using two different devices: Penguin RFA and Osstell IDX.

RESULTS

Primary stability was superior in the cortical bone in both torque and RFA. In the block containing cortical bone, group A implants obtained a greater insertion torque than did group B. The insertion torque was lesser in the bone lacking cortex. Regarding the ISQ of the implants, group A presented higher values in the block with cortical bone, but the values were lower in the block without cortical bone. There were no significant differences between the values obtained from the Osstell IDX and Penguin RFA.

CONCLUSIONS

The presence of cortical bone positively influences the primary stability of dental implants. The design of the implant also has a statistically significant influence on implant primary stability, although the impact depends on whether there is coronal cerclage or not. There were no statistically significant differences in the implant stability measurements obtained by two different devices. Implant stability, resonance frequency analysis, torque, osstell, penguin, cortical.

摘要

背景

本研究旨在评估种植体设计及皮质骨的存在对初期稳定性的影响,并分析两种不同共振频率分析(RFA)设备所测得的稳定性测量值之间的差异。

材料与方法

共使用了两种不同型号的80颗克洛克纳种植体[40颗Essential Cone种植体(A组)和40颗Vega种植体(B组)]。将种植体植入两个模拟上颌骨力学性能的聚氨酯块中。一个块体有一层皮质骨,另一个块体没有。使用两种不同的设备:Penguin RFA和Osstell IDX,通过植入扭矩和RFA测量所有种植体的初期稳定性。

结果

在扭矩和RFA方面,皮质骨中的初期稳定性均更佳。在含有皮质骨的块体中,A组种植体获得的植入扭矩大于B组。在缺乏皮质的骨中,植入扭矩较小。关于种植体的ISQ,A组在有皮质骨的块体中呈现出更高的值,但在没有皮质骨的块体中值较低。从Osstell IDX和Penguin RFA获得的值之间没有显著差异。

结论

皮质骨的存在对牙种植体的初期稳定性有积极影响。种植体的设计对种植体初期稳定性也有统计学上的显著影响,尽管其影响取决于是否有冠部环扎。两种不同设备所测得的种植体稳定性测量值没有统计学上的显著差异。种植体稳定性、共振频率分析、扭矩、Osstell、Penguin、皮质。