Horemans Herwin, Kooijmans Hedwig, van den Berg-Emons Rita, Bussmann Hans
Erasmus MC, Rotterdam, Netherlands.
J Rehabil Assist Technol Eng. 2020 Mar 16;7:2055668319890535. doi: 10.1177/2055668319890535. eCollection 2020 Jan-Dec.
To set up and guide interventions with the aim to increase physical activity and lower sedentary behavior valid monitoring of physical behavior is essential. The aim of this study was to evaluate the validity of the single-unit Activ8 activity monitor to classify several body postures and movements.
Twelve healthy adults performed a series of activities, representative for everyday life, according to a standard protocol. Activ8 was both worn in the trouser pocket (prescribed location) and fixated to the front of the thigh. Activities were video recorded and analyzed thereafter. Postures and movements that were analyzed were lying/sitting, standing, walking, cycling, and running.
The agreement between Activ8 output and video analysis was 89.7% (inter-subject range: 66.0 to 96.6%) for the pocket location and 91.9% (range 85.5 to 95.1%) for the thigh location. Sensitivity and positive predictive value scores for both locations were all above 80%, except for standing (69% or higher). Differences in classified duration of separate postures and movements were within 20% for walking, sitting and running.
The Activ8 is a valid instrument to quantify a defined set of body postures and movements. Because of the smaller time difference, the thigh location is preferred for research purposes.
为了建立并指导旨在增加身体活动和减少久坐行为的干预措施,对身体行为进行有效的监测至关重要。本研究的目的是评估单单元Activ8活动监测器对几种身体姿势和动作进行分类的有效性。
12名健康成年人按照标准方案进行了一系列代表日常生活的活动。Activ8既佩戴在裤兜(规定位置),也固定在大腿前部。活动之后进行了视频录制和分析。分析的姿势和动作包括躺/坐、站立、行走、骑自行车和跑步。
对于裤兜位置,Activ8输出与视频分析之间的一致性为89.7%(个体间范围:66.0%至96.6%),对于大腿位置,一致性为91.9%(范围85.5%至95.1%)。除站立(69%或更高)外,两个位置的敏感性和阳性预测值分数均高于80%。对于行走、坐立和跑步,各个姿势和动作分类持续时间的差异在20%以内。
Activ8是一种用于量化一组特定身体姿势和动作的有效仪器。由于时间差异较小,研究目的首选大腿位置。