Pan Xiaojing, Maram Jyotsna, Nittala Muneeswar G, Francis Brian A, Chopra Vikas, Sadda Srinivas R
Qingdao Eye Hospital, Shandong Eye Institute, Shandong First Medical University & Shandong Academy of Medical Sciences, Qingdao, Shandong, People's Republic of China.
Doheny Image Reading Center, Doheny Eye Institute, University of California Los Angeles, Los Angeles, CA, USA.
Graefes Arch Clin Exp Ophthalmol. 2020 Jul;258(7):1475-1481. doi: 10.1007/s00417-020-04648-1. Epub 2020 Mar 26.
To compare the reproducibility and agreement of anterior chamber angle (ACA) parameters and metrics obtained by four different anterior segment-optical coherence tomography (AS-OCT) devices.
In this prospective study, 30 eyes from 15 normal subjects underwent anterior segment angle scanning using the Spectralis, Cirrus, and Optovue spectral domain optical coherence tomography (SD-OCT), as well as the Visante time-domain optical coherence tomography (TD-OCT). For each eye, the scan line was performed perpendicularly on the inferior (270°) angle, and the inferior ACA image was acquired 2 times. Inter-instrument and intra-instrument, as well as inter-observer and intra-observer reproducibility of anterior chamber angle metrics, Schwalbe's line (SL) to scleral spur (SS) distance (TM-Span), angle opening distance (AOD), and trabecular iris space area (TISA) measurements, were evaluated by intraclass correlation coefficients (ICCs) and Bland-Altman plots with limits of agreement (LoA).
For this cohort of 30 eyes of 15 normal subjects, the mean TM-Span, AOD, and TISA were 0.966 ± 0.198 mm, 0.750 ± 0.205 mm, and 0.286 ± 0.090 mm from the Spectralis; 0.929 ± 0.113 mm, 0.717 ± 0.120 mm, and 0.267 ± 0.095 mm from the Cirrus; 0.923 ± 0.191 mm, 0.683 ± 0.161 mm, and 0.265 ± 0.072 mm from the Optovue; and 0.970 ± 0.070 mm, 0.705 ± 0.150 mm, and 0.279 ± 0.065 mm from the Visante. The intra-instrument (ICCs > 0.838), intra-grader (ICCs > 0.910), and inter-grader (ICCs > 0.869) agreement were good. Agreement between the four instruments was also good with ICCs from 0.901 to 0.967 for TM-Span, 0.887 to 0.941 for AOD, and 0.923 to 0.961 for TISA.
Consistent and reproducible ACA measurements could be obtained from multiple AS-OCT devices including both SD- and TD-OCT instruments. These findings have relevance when considering multiple imaging devices in future studies.
比较四种不同的眼前节光学相干断层扫描(AS-OCT)设备所获得的前房角(ACA)参数和测量指标的可重复性及一致性。
在这项前瞻性研究中,对15名正常受试者的30只眼睛使用Spectralis、Cirrus和Optovue光谱域光学相干断层扫描(SD-OCT)以及Visante时域光学相干断层扫描(TD-OCT)进行眼前节角扫描。对于每只眼睛,扫描线垂直于下方(270°)角进行,获取下方ACA图像2次。通过组内相关系数(ICC)和带有一致性界限(LoA)的Bland-Altman图,评估前房角测量指标、Schwalbe线(SL)至巩膜突(SS)距离(TM-Span)、房角开放距离(AOD)和小梁虹膜空间面积(TISA)测量在仪器间、仪器内以及观察者间和观察者内的可重复性。
对于这15名正常受试者的30只眼睛组成的队列,Spectralis测量的平均TM-Span、AOD和TISA分别为0.966±0.198mm、0.750±0.205mm和0.286±0.090mm;Cirrus测量的分别为0.929±0.113mm、0.717±0.120mm和0.267±0.095mm;Optovue测量的分别为0.923±0.191mm、0.683±0.161mm和0.265±0.072mm;Visante测量的分别为0.970±0.070mm、0.705±0.150mm和0.279±0.065mm。仪器内(ICC>0.838)、分级者内(ICC>0.910)和分级者间(ICC>0.869)的一致性良好。四种仪器之间的一致性也良好,TM-Span的ICC为0.901至0.967,AOD的ICC为0.887至0.941,TISA的ICC为0.923至0.961。