Imholz B P, Wieling W, Langewouters G J, van Montfrans G A
Department of Internal Medicine, Academic Medical Centre, Amsterdam, The Netherlands.
Clin Auton Res. 1991 Mar;1(1):43-53. doi: 10.1007/BF01826057.
The clinical utility for autonomic research purposes of the Finapres, a device for measuring finger arterial pressure continuously and noninvasively, was assessed by estimating its accuracy, precision and within-subject variability in 48 subjects, aged 18-65 years, in comparison with intraarterial brachial blood pressure. At differences of -3.5 +/- 12, -8.0 +/- 8 and -4.4 +/- 8 mmHg from simultaneous intrabrachial systolic, mean and diastolic pressure, the Finapres meets the Association for the Advancement of Medical Instruments (AAMI) accuracy requirements for systolic and diastolic, but not for mean pressure. The precision requirements were nearly met for mean and diastolic, but not for systolic pressure. These results compare to those of others under widely varying circumstances in anaesthetized and awake subjects, and are comparable to those published for the auscultatory technique. The within-subject precision of Finapres is high; the 95% confidence intervals are 3.4-4.5 mmHg for systolic, 1.5-2.0 mmHg for mean and 1.7-2.2 mmHg for diastolic. This makes the Finapres suitable for tracking changes in blood pressure. Four case studies are provided as examples of the value of the Finapres in the clinical laboratory.
通过与肱动脉内血压相比较,评估了用于自主神经研究的连续无创测量手指动脉压的设备Finapres在48名年龄在18至65岁的受试者中的准确性、精密度和个体内变异性,以此来评估其临床实用性。与同时测量的肱动脉收缩压、平均压和舒张压相差-3.5±12、-8.0±8和-4.4±8 mmHg时,Finapres符合美国医疗仪器促进协会(AAMI)对收缩压和舒张压的准确性要求,但不符合对平均压的要求。对于平均压和舒张压,几乎满足精密度要求,但对于收缩压则不满足。这些结果与在麻醉和清醒受试者的广泛不同情况下其他人的结果进行了比较,并且与听诊技术公布的结果相当。Finapres的个体内精密度很高;收缩压的95%置信区间为3.4 - 4.5 mmHg,平均压为1.5 - 2.0 mmHg,舒张压为1.7 - 2.2 mmHg。这使得Finapres适用于追踪血压变化。提供了四个案例研究作为Finapres在临床实验室中的价值示例。