To investigate interdevice agreement among corneal topography/ray-tracing aberrometry (iTrace), partial coherence interferometry (IOLMaster), and Scheimpflug imaging (Pentacam) for the measurement of corneal astigmatism.
The analysis included 90 eyes of 90 subjects without ocular disease. The main outcome measures were corneal cylinder power and axis of astigmatism. All corneal astigmatism measurements were converted to two perpendicular components by using vector analysis. Interdevice agreement was assessed using Bland-Altman analysis, paired sample -test, and one-way analysis of variance.
No significant interdevice difference existed in the astigmatism magnitude, cardinal component, and oblique component (all > 0.05). On comparing iTrace wavefront and simulated keratometry (SimK) astigmatism, significant differences were observed in the astigmatism magnitude and oblique component (both < 0.01), but not in the cardinal component (=0.687). On comparing Pentacam pupil 3 mm and corneal vertex 3 mm axial astigmatism, significant difference was observed in the astigmatism magnitude ( < 0.001), but not in the cardinal and oblique components (both > 0.05).
The iTrace, IOLMaster, and Pentacam devices could be used interchangeably for corneal astigmatism measurement. However, the measurement difference in iTrace wavefront and SimK astigmatism and Pentacam pupil 3 mm and vertex 3 mm axial astigmatism should be considered in clinic practice.
研究角膜地形图/光线追踪像差仪(iTrace)、部分相干干涉仪(IOLMaster)和Scheimpflug成像仪(Pentacam)在测量角膜散光方面的设备间一致性。
分析纳入90例无眼部疾病受试者的90只眼。主要观察指标为角膜柱镜度数和散光轴。通过矢量分析将所有角膜散光测量值转换为两个垂直分量。采用Bland-Altman分析、配对样本t检验和单因素方差分析评估设备间一致性。
在散光量、主要分量和斜向分量方面,各设备间无显著差异(均P>0.05)。比较iTrace波前像差和模拟角膜曲率计(SimK)散光时,在散光量和斜向分量上观察到显著差异(均P<0.01),但在主要分量上无差异(P=0.687)。比较Pentacam 3mm瞳孔和角膜顶点3mm轴向散光时,在散光量上观察到显著差异(P<0.001),但在主要分量和斜向分量上无差异(均P>0.05)。
iTrace、IOLMaster和Pentacam设备在角膜散光测量中可互换使用。然而,在临床实践中应考虑iTrace波前像差与SimK散光以及Pentacam 3mm瞳孔和顶点3mm轴向散光的测量差异。