眼前节角膜像差测量、角膜曲率计测量及瞳孔大小与眼前节分析系统和光线追踪像差仪的比较

Comparison of Anterior Corneal Aberrometry, Keratometry and Pupil Size with Scheimpflug Tomography and Ray Tracing Aberrometer.

作者信息

Ashena Zahra, Gallagher Sean, Naveed Hasan, Spalton David J, Nanavaty Mayank A

机构信息

Sussex Eye Hospital, University Hospitals Sussex NHS Foundation Trust, Brighton BN2 5BF, UK.

Maidstone Hospital, Maidstone and Tunbridge Wells NHS Trust, Maidstone TN2 4QJ, UK.

出版信息

Vision (Basel). 2022 Mar 18;6(1):18. doi: 10.3390/vision6010018.

Abstract

This study aimed to assess the anterior corneal wavefront aberrations, keratometry, astigmatism vectors and pupil size between Pentacam HR® (Oculus Optikgeraete GmbH, Wetzlar, Germany) and iTrace® (Tracey Technologies Corp., Houston, TX, USA). In this observational study, 100 eyes (50 healthy volunteers) were scanned in mesopic light condition with a Pentacam HR® and iTrace®. Anterior corneal aberrations (spherical aberration (Z40), vertical coma (Z3 − 1), horizontal coma (Z3 + 1)), keratometry in the flattest (K1) and steepest meridian (K2), mean astigmatism, astigmatic vectors (J0 and J45), and pupil size were measured. We found a significant difference in Z40 (Pentacam®: +0.30 ± 0.11 µm and iTrace®: −0.03 µm ± 0.05 µm; p < 0.01) with no correlation between the devices (r = −0.12, p = 0.22). The devices were in complete agreement for Z3 − 1 (p = 0.78) and Z3 + 1 (p = 0.39), with significant correlation between the machines (r = −0.38, p < 0.01 and r = −0.6, p < 0.01). There was no difference in K1, K2 and mean astigmatism. J0 was negative with both devices (against-the-rule astigmatism), but there was no correlation. J45 was negative with the Pentacam HR® (more myopic oblique astigmatism) but significantly correlated between the devices. Pupil size was smaller with Pentacam HR® (p < 0.01). In summary, these devices cannot be used interchangeably. Corneal Z40 was significantly different with more negative Z40 with iTrace® compared to Pentacam HR®. iTrace® operates with lower illumination, giving larger pupil size than Pentacam HR®, which uses intense blue light during measurement. No correlation was found for J0. Pentacam HR® had a trend to record more negative J45 (myopic oblique astigmatism).

摘要

本研究旨在评估Pentacam HR®(德国韦茨拉尔奥culus Optikgeraete有限公司)和iTrace®(美国得克萨斯州休斯顿Tracey Technologies公司)之间的角膜前表面波前像差、角膜曲率、散光矢量和瞳孔大小。在这项观察性研究中,100只眼睛(50名健康志愿者)在中度照明条件下使用Pentacam HR®和iTrace®进行扫描。测量了角膜前表面像差(球差(Z40)、垂直彗差(Z3 - 1)、水平彗差(Z3 + 1))、最平子午线(K1)和最陡子午线(K2)的角膜曲率、平均散光、散光矢量(J0和J45)以及瞳孔大小。我们发现Z40存在显著差异(Pentacam®:+0.30±0.11 µm,iTrace®:−0.03 µm±0.05 µm;p<0.01),且两种设备之间无相关性(r = -0.12,p = 0.22)。两种设备在Z3 - 1(p = 0.78)和Z3 + 1(p = 0.39)方面完全一致,且两台机器之间存在显著相关性(r = -0.38,p<0.01和r = -0.6,p<0.01)。K1、K2和平均散光无差异。两种设备测得的J0均为负值(逆规散光),但无相关性。Pentacam HR®测得的J45为负值(更多近视斜向散光),但两种设备之间存在显著相关性。Pentacam HR®测得的瞳孔大小较小(p<0.01)。总之,这些设备不能互换使用。与Pentacam HR®相比,角膜Z40存在显著差异,iTrace®测得的Z40更负。iTrace®在较低照明条件下运行,其测得的瞳孔大小比Pentacam HR®大,Pentacam HR®在测量过程中使用强蓝光。未发现J0存在相关性。Pentacam HR®有记录到更多负J45(近视斜向散光)的趋势。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/a9dc/8951144/70ef4ae81d8e/vision-06-00018-g001.jpg