Yan Rengna, Li Huiqin, Kong Xiaocen, Zhai Xiaofang, Chen Maoyuan, Sun Yixuan, Ye Lei, Su Xiaofei, Ma Jianhua
Department of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing 210012, China.
National Heart Research Institute Singapore, National Heart Centre Singapore, Singapore.
Int J Endocrinol. 2020 Jan 4;2020:5947680. doi: 10.1155/2020/5947680. eCollection 2020.
The purpose of this study was to investigate the accuracy of the continuously stored data from the Abbott FreeStyle Libre flash glucose monitoring (FGM) system in Chinese diabetes patients during standard meal tests when glucose concentrations were rapidly changing. . Interstitial glucose levels were monitored for 14 days in 26 insulin-treated patients with type 2 diabetes using the FGM system. Standard meal tests were conducted to induce large glucose swings. Venous blood glucose (VBG) was tested at 0, 30, 60, and 120 min after standard meal tests in one middle day of the first and second weeks, respectively. The corresponding sensor glucose values were obtained from interpolating continuously stored data points. Assessment of accuracy was according to recent consensus recommendations with median absolute relative difference (MARD) and Clarke and Parkes error grid analysis (CEG and PEG).
Among 208 paired sensor-reference values, 100% were falling within zones A and B of the Clarke and Parkes error grid analysis. The overall MARD was 10.7% (SD, 7.8%). Weighted least squares regression analysis resulted in high agreement between the FGM sensor glucose and VBG readings. The overall MTT results showed that FGM was lower than actual VBG, with MAD of 22.1 mg/dL (1.2 mmol/L). At VBG rates of change of -1 to 0, 0 to 1, 1 to 2, and 2 to 3 mg/dl/min, MARD results were 11.4% (SD, 8.7%), 9.4% (SD, 6.5%), 9.9% (SD, 7.5%), and 9.5% (SD, 7.7%). At rapidly changing VBG concentrations (>3 mg/dl/min), MARD increased to 19.0%, which was significantly higher than slow changing BG groups.
Continuously stored interstitial glucose measurements with the FGM system were found to be acceptable to evaluate VBG in terms of clinical decision during standard meal tests. The continuously stored data from the FGM system appeared to underestimate venous glucose and performed less well during rapid glucose changes.
本研究的目的是调查在中国糖尿病患者进行标准餐试验期间,当血糖浓度快速变化时,雅培瞬感扫描式葡萄糖监测(FGM)系统连续存储数据的准确性。使用FGM系统对26例接受胰岛素治疗的2型糖尿病患者的组织间液葡萄糖水平进行了14天的监测。进行标准餐试验以诱导血糖大幅波动。在第一周和第二周的中间一天,分别在标准餐试验后的0、30、60和120分钟检测静脉血糖(VBG)。通过对连续存储的数据点进行插值获得相应的传感器葡萄糖值。根据最近的共识建议,采用中位数绝对相对差异(MARD)以及克拉克和帕克斯误差网格分析(CEG和PEG)进行准确性评估。
在208对传感器 - 参考值中,100%落在克拉克和帕克斯误差网格分析的A区和B区。总体MARD为10.7%(标准差,7.8%)。加权最小二乘回归分析结果显示FGM传感器葡萄糖读数与VBG读数高度一致。总体平均绝对差值(MTT)结果显示,FGM低于实际VBG,平均绝对差值(MAD)为22.1mg/dL(1.2mmol/L)。在VBG变化率为 -1至0、0至1、1至2和2至3mg/dl/分钟时,MARD结果分别为11.4%(标准差,8.7%)、9.4%(标准差,6.5%)、9.9%(标准差,7.5%)和9.5%(标准差,7.7%)。在VBG浓度快速变化(>3mg/dl/分钟)时,MARD增加到19.0%,显著高于血糖缓慢变化组。
发现在标准餐试验期间,就临床决策而言,使用FGM系统连续存储的组织间液葡萄糖测量值可用于评估VBG。FGM系统连续存储的数据似乎低估了静脉血糖,并且在血糖快速变化期间表现较差。