Automated blood pressure (BP) monitor is widely used to assess the blood pressure (BP) of the study subjects in community-based researches. This study aims at the detection of hypertension by automated BP device and examines the concordance and validity between automated and mercury-free LED BP devices.
This cross-sectional study was conducted in the tribal state of Jharkhand in India from January 2017 to June 2017. A total of 300 study participants aged more than 18 years were enrolled in this study. BP of the patients in the sitting position was measured three times each by automated device and mercury-free LED BP device. The different sets of readings were assessed by concordance correlation coefficient (CCC) and other validity measures.
The CCC for systolic and diastolic BP measured by automated and mercury-free LED BP is 0.88 and 0.85, respectively. The mean difference between systolic and diastolic BP by both the instruments is statistically insignificant ( > 0.05). The sensitivity, specificity, PPV, NPV, and accuracy of automated BP devices to predict hypertension is 96.61%, 92.21%, 75%, 99%, and 93%. The area under ROC for systolic and diastolic BP is 0.984 and 0.97, respectively with values < 0.0001 in both the cases.
This study concluded that the overall automated BP machine has fair degree of agreement (CCC) with a manual BP device. The validity of this monitor to screen hypertension may also be considered in field settings.
在基于社区的研究中,自动血压监测仪被广泛用于评估研究对象的血压。本研究旨在通过自动血压设备检测高血压,并检验自动血压设备与无汞LED血压设备之间的一致性和有效性。
本横断面研究于2017年1月至2017年6月在印度贾坎德邦部落地区进行。本研究共纳入300名年龄超过18岁的研究参与者。通过自动设备和无汞LED血压设备分别对患者坐位时的血压进行三次测量。通过一致性相关系数(CCC)和其他有效性指标对不同组的读数进行评估。
自动血压设备和无汞LED血压设备测量的收缩压和舒张压的CCC分别为0.88和0.85。两种仪器测量的收缩压和舒张压的平均差异无统计学意义(>0.05)。自动血压设备预测高血压的敏感性、特异性、阳性预测值、阴性预测值和准确性分别为96.61%、92.21%、75%、99%和93%。收缩压和舒张压的ROC曲线下面积分别为0.984和0.97,两种情况下P值均<0.0001。
本研究得出结论,总体而言自动血压仪与手动血压设备具有相当程度的一致性(CCC)。在现场环境中也可考虑该监测仪筛查高血压的有效性。