The authors aimed to quantify end-digit and threshold biases in blood pressure (BP) measurement with manual and digital sphygmomanometers. In a 3-year follow-up, end-digit and threshold biases were investigated and a new index, called the deviation index, was used to quantify measurement bias. The distribution of systolic and diastolic BPs became close to normal after implementation of digital sphygmomanometers. The appearance of zero end digits decreased from 97% to 30% (P<.0001). The deviation index decreased from 97% to 20% (P<.0001). Mean systolic and diastolic BPs increased immediately after implementation of automated sphygmomanometers (124.22±0.83 vs 132.90±0.78 and 74.38±0.50 vs 80.43±0.51, respectively; P<.0001 for both) but showed a linear decreasing trend during follow-up (systolic -3.59 mm Hg per year; 95% confidence interval, -5.57 to -1.61 [P<.0001]; diastolic: -2.52 mm Hg per year; 95% confidence interval, -3.78 to -1.26 [P<.0001]). Threshold bias decreased from 12.94% to 6.68% (P<.0001). Replacing manual sphygmomanometers with digital devices decreased end-digit and threshold biases in BP measurement. The deviation index can be used to quantify the magnitude of measurement bias.
作者旨在量化使用手动和数字血压计测量血压(BP)时的尾数和阈值偏差。在为期3年的随访中,研究了尾数和阈值偏差,并使用一种名为偏差指数的新指标来量化测量偏差。使用数字血压计后,收缩压和舒张压的分布变得接近正态分布。尾数为零的情况从97%降至30%(P<0.0001)。偏差指数从97%降至20%(P<0.0001)。使用自动血压计后,收缩压和舒张压的平均值立即升高(分别为124.22±0.83 vs 132.90±0.78以及74.38±0.50 vs 80.43±0.51;两者P均<0.0001),但在随访期间呈线性下降趋势(收缩压每年下降3.59 mmHg;95%置信区间为-5.57至-1.61[P<0.0001];舒张压:每年下降2.52 mmHg;95%置信区间为-3.78至-1.26[P<0.0001])。阈值偏差从12.94%降至6.68%(P<0.0001)。用数字设备取代手动血压计可减少血压测量中的尾数和阈值偏差。偏差指数可用于量化测量偏差的大小。