Yánez-Jácome Gabriela S, Romero-Estévez David, Navarrete Hugo, Simbaña-Farinango Karina, Vélez-Terreros Pamela Y
Centro de Estudios Aplicados en Química, Pontificia Universidad Católica del Ecuador, Quito 17012184, Ecuador.
Methods Protoc. 2020 Jun 23;3(2):45. doi: 10.3390/mps3020045.
Several microwave-assisted digestion methods were tested at the Centro de Estudios Aplicados en Química laboratory in Quito, Ecuador, to determine the accuracy and performance efficiency of the mineralization process for the determination of total mercury in fish tissue by cold vapor atomic fluorescence spectrophotometry. The use of MARSEasyPrep high-pressure vessels, low amounts of reagents (1 cm HNO, 1 cm HO, and 1 cm HClO), an irradiation temperature of 210 °C, and 35 min of mineralization time resulted in accurate performance, with recoveries of certified reference material DORM-4 between 90.1% and 105.8%. This is better than the Association of Official Analytical Chemists 2015.01 method, which has a reported accuracy of 81%. The repeatability precision and intermediate precision were established at three concentration levels (0.167, 0.500, and 0.833 mg·kg) and expressed as the percentage of the relative standard deviation ranging from 1.5% to 3.0% and 1.7% to 4.2%, respectively. Further, the method was satisfactorily applied to analyze fortified samples of tilapia (), with recoveries ranging from 98.3% to 104.3%. The instrumental limits of detection and quantification were 0.118 µg·dm and 0.394 µg·dm, respectively.
在厄瓜多尔基多的应用化学研究中心实验室测试了几种微波辅助消解方法,以确定通过冷蒸气原子荧光分光光度法测定鱼组织中总汞的矿化过程的准确性和性能效率。使用MARSEasyPrep高压容器、少量试剂(1厘米硝酸、1厘米过氧化氢和1厘米高氯酸)、210℃的辐照温度和35分钟的矿化时间,获得了准确的结果,有证标准物质DORM-4的回收率在90.1%至105.8%之间。这比美国公职分析化学师协会2015.01方法更好,该方法报告的准确度为81%。在三个浓度水平(0.167、0.500和0.833毫克·千克)下确定了重复性精密度和中间精密度,分别以相对标准偏差百分比表示,范围为1.5%至3.0%和1.7%至4.2%。此外,该方法成功应用于分析罗非鱼的加标样品,回收率在98.3%至104.3%之间。仪器的检测限和定量限分别为0.118微克·立方分米和0.394微克·立方分米。