Department of Clinical Laboratory, Renmin Hospital of Wuhan University, Wuhan, 430060, China.
Curr Med Sci. 2020 Aug;40(4):614-617. doi: 10.1007/s11596-020-2224-y. Epub 2020 Aug 29.
The novel Coronavirus SARS-CoV-2 caused an outbreak of pneumonia in Wuhan, Hubei province of China in January 2020. This study aims to investigate the effects of different temperature and time durations of virus inactivation on the results of PCR testing for SARS-CoV-2. Twelve patients at the Renmin Hospital of Wuhan University suspected of being infected with SARS-CoV-2 were selected on February 13, 2020 and throat swabs were taken. The swabs were stored at room temperature (20-25°C), then divided into aliquots and subjected to different temperature for different periods in order to inactivate the viruses (56°C for 30, 45, 60 min; 65, 70, 80°C for 10, 15, 20 min). Control aliquots were stored at room temperature for 60 min. Then all aliquots were tested in a real-time fluorescence PCR using primers against SARS-CoV-2. Regardless of inactivation temperature and time, 7 of 12 cases (58.3%) tested were positive for SARS-CoV-2 by PCR, and cycle threshold values were similar. These results suggest that virus inactivation parameters exert minimal influence on PCR test results. Inactivation at 65°C for 10 min may be sufficient to ensure safe, reliable testing.
新型冠状病毒 SARS-CoV-2 于 2020 年 1 月在中国湖北省武汉市引发了肺炎疫情。本研究旨在探讨不同病毒灭活温度和时间对 SARS-CoV-2 PCR 检测结果的影响。2020 年 2 月 13 日,选取了 12 名在武汉大学人民医院疑似感染 SARS-CoV-2 的患者,并采集了咽拭子。将拭子在室温(20-25°C)下保存,然后分成等分试样,并在不同温度下进行不同时间的处理以灭活病毒(56°C 30、45、60 分钟;65、70、80°C 10、15、20 分钟)。对照等分试样在室温下保存 60 分钟。然后,使用针对 SARS-CoV-2 的引物通过实时荧光 PCR 对所有等分试样进行检测。无论灭活温度和时间如何,12 例中有 7 例(58.3%)通过 PCR 检测对 SARS-CoV-2 呈阳性,且循环阈值相似。这些结果表明,病毒灭活参数对 PCR 检测结果的影响很小。65°C 灭活 10 分钟可能足以确保安全、可靠的检测。