右侧Fontan辅助装置的计算建模:不同患者解剖结构和插管方式下的有效性

Computational modeling of a right-sided Fontan assist device: Effectiveness across patient anatomies and cannulations.

作者信息

Trusty Phillip M, Alan Wei Zhenglun, Fogel Mark A, Maher Kevin, Deshpande Shriprasad R, Yoganathan Ajit P

机构信息

Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology, Atlanta, GA, United States.

Division of Cardiology, Children's Hospital of Philadelphia, Philadelphia, PA, United States.

出版信息

J Biomech. 2020 Aug 26;109:109917. doi: 10.1016/j.jbiomech.2020.109917. Epub 2020 Jul 2.

Abstract

The use of mechanical circulatory support for failing Fontan patients is an area of growing interest, as the increased life expectancy of these patients continues to be accompanied by numerous end-organ complications. In vitro work has shown positive results using the CentriMag device for right-sided Fontan support, however the generalizability across various patient anatomies and cannulations is unknown. Computational simulations are first validated against in vitro modeling, then used to assess generalizability and further explore hemodynamic metrics including relative pressure changes, hepatic flow distribution, wall shear stress and power added. Computational modeling matched previous in vitro work very well, with vessel flow rates and relative average pressure change each within 1%. Positive results were seen across all patient anatomies and cannulations. On average, pressure from the vena cava to pulmonary arteries increased by 5.4 mmHg corresponding to 32 mW of power added. Hepatic flow distribution and wall shear stress were within acceptable ranges, with an average hepatic flow distribution of 47% and all patients showing ≤ 1% of the total Fontan connection surface area at a wall shear stress above 150 Pa. The positive results previously seen using CentriMag as a right-sided Fontan support device were found to be repeatable across multiple patient anatomies and cannulations. While animal models and eventual patient studies will provide further insight into the efficacy of this support strategy, our findings here suggest this method may reproduce right heart function.

摘要

对于衰竭的Fontan手术患者使用机械循环支持是一个日益受到关注的领域,因为这些患者预期寿命的延长仍伴随着众多终末器官并发症。体外研究表明,使用CentriMag装置进行右侧Fontan支持取得了积极成果,然而,其在各种患者解剖结构和插管方式中的通用性尚不清楚。首先将计算模拟与体外模型进行验证,然后用于评估通用性并进一步探索血流动力学指标,包括相对压力变化、肝血流分布、壁面剪应力和额外功率。计算模型与先前的体外研究结果非常吻合,血管流速和相对平均压力变化均在1%以内。在所有患者解剖结构和插管方式中均观察到了积极结果。平均而言,从腔静脉到肺动脉的压力增加了5.4 mmHg,相应地增加了32 mW的功率。肝血流分布和壁面剪应力在可接受范围内,平均肝血流分布为47%,所有患者在壁面剪应力高于150 Pa时,Fontan连接总面积的比例均≤1%。先前使用CentriMag作为右侧Fontan支持装置所观察到的积极结果在多种患者解剖结构和插管方式中均可重复。虽然动物模型和最终的患者研究将进一步深入了解这种支持策略的疗效,但我们在此的研究结果表明,这种方法可能重现右心功能。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验