Trends Hear. 2020 Jan-Dec;24:2331216520948974. doi: 10.1177/2331216520948974.
Automatic gain control (AGC) compresses the wide dynamic range of sounds to the narrow dynamic range of hearing-impaired listeners. Setting AGC parameters (time constants and knee points) is an important part of the fitting of hearing devices. These parameters do not only influence overall loudness elicited by the hearing devices but can also affect the recognition of speech in noise. We investigated whether matching knee points and time constants of the AGC between the cochlear implant and the hearing aid of bimodal listeners would improve speech recognition in noise. We recruited 18 bimodal listeners and provided them all with the same cochlear-implant processor and hearing aid. We compared the matched AGCs with the default device settings with mismatched AGCs. As a baseline, we also included a condition with the mismatched AGCs of the participants' own devices. We tested speech recognition in quiet and in noise presented from different directions. The time constants affected outcomes in the monaural testing condition with the cochlear implant alone. There were no specific binaural performance differences between the two AGC settings. Therefore, the performance was mostly dependent on the monaural cochlear implant alone condition.
自动增益控制(AGC)将声音的宽动态范围压缩到听力受损听众的窄动态范围。设置 AGC 参数(时间常数和拐点)是听力设备适配的重要组成部分。这些参数不仅会影响听力设备产生的整体响度,还会影响噪声中的语音识别。我们研究了在双耳模式的听众中,匹配人工耳蜗和助听器的 AGC 拐点和时间常数是否会提高噪声中的语音识别能力。我们招募了 18 名双耳模式的听众,并为他们提供了相同的人工耳蜗处理器和助听器。我们将匹配的 AGC 与默认设备设置和不匹配的 AGC 进行了比较。作为基线,我们还包括了参与者自己设备的不匹配 AGC 的条件。我们在安静和来自不同方向的噪声中测试了语音识别。在单独使用人工耳蜗的单耳测试条件下,时间常数会影响结果。两种 AGC 设置之间没有特定的双耳性能差异。因此,性能主要取决于单独使用人工耳蜗的单耳条件。