Department of Urology, The Seventh Medical Center of Chinese PLA General Hospital, No 5 Nanmencang, Beijing, 100700, China.
Department of Urology, The Sixth Medical Center of Chinese PLA General Hospital, Beijing, 100048, China.
Int Urol Nephrol. 2021 Mar;53(3):415-419. doi: 10.1007/s11255-020-02665-x. Epub 2020 Sep 28.
To evaluate calyceal irrigation fluid temperature changes during flexible ureteroscopic Ho:YAG laser lithotripsy.
Between May 2019 and January 2020, patients with kidney stones undergoing flexible ureteroscopic Ho:YAG laser lithotripsy were enrolled. A K-type thermocouple was applied for intraoperative temperature measurement. Laser was activated at different power (1 J/20 Hz and 0.5 J/20 Hz) and irrigation (0 ml/min, 15 ml/min and 30 ml/min) settings, temperature-time curve was drawn and time needed to reach 43 °C without irrigation was documented.
Thirty-two patients were enrolled in our study. The temperature-time curve revealed a quick temperature increase followed by a plateau. With 15 ml/min or 30 ml/min irrigation, 43 °C was not reached after 60 s laser activation at both 1 J/20 Hz and 0.5 J/20 Hz. At the power setting of 1 J/20 Hz and irrigation flow rate of 15 ml/min, the temperature rise was significantly higher than other groups. Without irrigation, the time needed to reach 43 °C at 1 J/20 Hz was significantly shorter than that at 0.5 J/20 Hz (8.84 ± 1.41 s vs. 13.71 ± 1.53 s).
Ho:YAG laser lithotripsy can induce significant temperature rise in calyceal fluid. With sufficient irrigation, temperatures can be limited so that a toxic thermal dose is not reached, when irrigation is closed, the temperature increased sharply and reached 43 °C in a few seconds.
评估软性输尿管镜钬激光碎石术中肾盂冲洗液温度的变化。
2019 年 5 月至 2020 年 1 月,纳入接受软性输尿管镜钬激光碎石术的肾结石患者。术中采用 K 型热电偶进行温度测量。激光以不同功率(1 J/20 Hz 和 0.5 J/20 Hz)和冲洗(0 ml/min、15 ml/min 和 30 ml/min)设置激活,绘制温度-时间曲线,并记录无冲洗时达到 43°C 所需的时间。
本研究共纳入 32 例患者。温度-时间曲线显示温度迅速升高后达到平台期。在 15 ml/min 或 30 ml/min 冲洗下,在 1 J/20 Hz 和 0.5 J/20 Hz 时,激光激活 60 s 后未达到 43°C。在 1 J/20 Hz 功率设置和 15 ml/min 冲洗流量下,温度升高明显高于其他组。无冲洗时,1 J/20 Hz 达到 43°C 的时间明显短于 0.5 J/20 Hz(8.84 ± 1.41 s 比 13.71 ± 1.53 s)。
钬激光碎石术可引起肾盂冲洗液显著升温。当冲洗充足时,温度可以被限制,以避免达到毒性热剂量;当冲洗关闭时,温度会急剧升高,并在几秒钟内达到 43°C。