Department of Urology, Gazi State Hospital, Ilkadım, Samsun, Turkey.
Department of Urology, Faculty of Medicine, Ondokuz Mayıs University, Samsun, Turkey.
Urolithiasis. 2024 Jun 13;52(1):86. doi: 10.1007/s00240-024-01592-1.
Ho: YAG laser lithotripsy is widely used for urinary stone treatment, but concerns persist regarding its thermal effects on renal tissues. This study aimed to monitor intrarenal temperature changes during kidney stone treatment using retrograde intrarenal surgery with Ho: YAG laser. Fifteen patients were enrolled. Various laser power settings (0.8 J/10 Hz, 1.2 J/12 Hz) and irrigation modes (10 cc/min, 15 cc/min, 20 cc/min, gravity irrigation, and manual pump irrigation) were used. A sterile thermal probe was attached to a flexible ureterorenoscope and delivered into the calyceal system via the ureteral access sheath. Temperature changes were recorded with a T-type thermal probe with ± 0.1 °C accuracy. Laser power significantly influenced mean temperature, with a 4.981 °C difference between 14 W and 8 W laser power (p < 0.001). The mean temperature was 2.075 °C higher with gravity irrigation and 2.828 °C lower with manual pump irrigation (p = 0.038 and p = 0.005, respectively). Body mass index, laser power, irrigation model, and operator duty cycle explained 49.5% of mean temperature variability (Adj. R = 0.495). Laser power and operator duty cycle positively impacted mean temperature, while body mass index and specific irrigation models affected it negatively. Laser power and irrigation rate are critical for intrarenal temperature during Ho: YAG laser lithotripsy. Optimal settings and irrigation strategies are vital for minimizing thermal injury risk. This study underscores the need for ongoing research to understand and mitigate thermal effects during laser lithotripsy.
钬激光碎石术被广泛应用于治疗尿路结石,但仍存在对其在肾组织中产生的热效应的担忧。本研究旨在通过逆行性肾内手术监测钬:YAG 激光碎石术中肾内温度的变化。共纳入 15 例患者。采用不同的激光功率设置(0.8J/10Hz,1.2J/12Hz)和冲洗模式(10cc/min,15cc/min,20cc/min,重力冲洗和手动泵冲洗)。将无菌热探头连接到柔性输尿管镜上,并通过输尿管进入鞘输送到肾盏系统。使用 T 型热敏探头以±0.1°C 的精度记录温度变化。激光功率对平均温度有显著影响,14W 和 8W 激光功率之间的平均温度差异为 4.981°C(p<0.001)。重力冲洗时平均温度升高 2.075°C,手动泵冲洗时平均温度降低 2.828°C(p=0.038 和 p=0.005)。体重指数、激光功率、冲洗模式和操作器工作周期解释了平均温度变异性的 49.5%(调整后的 R=0.495)。激光功率和操作器工作周期对平均温度有正向影响,而体重指数和特定的冲洗模式对平均温度有负向影响。激光功率和冲洗速度是钬激光碎石术中肾内温度的关键因素。优化设置和冲洗策略对于最大限度地降低热损伤风险至关重要。本研究强调了需要进行持续研究,以了解和减轻激光碎石术中的热效应。