技术说明:从稀疏视图锥束CT数据进行4D锥束CT重建,用于笔形束扫描质子治疗(PBS-PT)中的日常运动评估。

Technical Note: 4D cone-beam CT reconstruction from sparse-view CBCT data for daily motion assessment in pencil beam scanned proton therapy (PBS-PT).

作者信息

den Otter Lydia A, Chen Kuanling, Janssens Guillaume, Meijers Arturs, Both Stefan, Langendijk Johannes A, Rosen Lane R, Wu Hsinshun T, Knopf Antje-Christin

机构信息

Department of Radiation Oncology, University Medical Center Groningen, University of Groningen, Groningen, 9713 GZ, The Netherlands.

Department of Radiation Oncology, Willis-Knighton Cancer Center, Shreveport, LA, USA.

出版信息

Med Phys. 2020 Dec;47(12):6381-6387. doi: 10.1002/mp.14521. Epub 2020 Oct 24.

Abstract

PURPOSE

The number of pencil beam scanned proton therapy (PBS-PT) facilities equipped with cone-beam computed tomography (CBCT) imaging treating thoracic indications is constantly rising. To enable daily internal motion monitoring during PBS-PT treatments of thoracic tumors, we assess the performance of Motion-Aware RecOnstructiOn method using Spatial and Temporal Regularization (MA-ROOSTER) four-dimensional CBCT (4DCBCT) reconstruction for sparse-view CBCT data and a realistic data set of patients treated with proton therapy.

METHODS

Daily CBCT projection data for nine non-small cell lung cancer (NSCLC) patients and one SCLC patient were acquired at a proton gantry system (IBA Proteus® One). Four-dimensional CBCT images were reconstructed applying the MA-ROOSTER and the conventional phase-correlated Feldkamp-Davis-Kress (PC-FDK) method. Image quality was assessed by visual inspection, contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and the structural similarity index measure (SSIM). Furthermore, gross tumor volume (GTV) centroid motion amplitudes were evaluated.

RESULTS

Image quality for the 4DCBCT reconstructions using MA-ROOSTER was superior to the PC-FDK reconstructions and close to FDK images (median CNR: 1.23 [PC-FDK], 1.98 [MA-ROOSTER], and 1.98 [FDK]; median SNR: 2.56 [PC-FDK], 4.76 [MA-ROOSTER], and 5.02 [FDK]; median SSIM: 0.18 [PC-FDK vs FDK], 0.31 [MA-ROOSTER vs FDK]). The improved image quality of MA-ROOSTER facilitated GTV contour warping and realistic motion monitoring for most of the reconstructions.

CONCLUSION

MA-ROOSTER based 4DCBCTs performed well in terms of image quality and appear to be promising for daily internal motion monitoring in PBS-PT treatments of (N)SCLC patients.

摘要

目的

配备锥形束计算机断层扫描(CBCT)成像用于治疗胸部疾病的笔形束扫描质子治疗(PBS-PT)设施数量不断增加。为了在胸部肿瘤的PBS-PT治疗期间实现每日内部运动监测,我们评估了使用空间和时间正则化的运动感知重建方法(MA-ROOSTER)对稀疏视图CBCT数据和质子治疗患者的真实数据集进行四维CBCT(4DCBCT)重建的性能。

方法

在质子龙门系统(IBA Proteus® One)上获取了9例非小细胞肺癌(NSCLC)患者和1例小细胞肺癌(SCLC)患者的每日CBCT投影数据。应用MA-ROOSTER和传统的相位相关菲尔德坎普-戴维斯-克雷斯(PC-FDK)方法重建四维CBCT图像。通过视觉检查、对比度噪声比(CNR)、信噪比(SNR)和结构相似性指数测量(SSIM)评估图像质量。此外,还评估了大体肿瘤体积(GTV)质心运动幅度。

结果

使用MA-ROOSTER进行的4DCBCT重建的图像质量优于PC-FDK重建,且接近FDK图像(中位数CNR:1.23 [PC-FDK],1.98 [MA-ROOSTER],1.98 [FDK];中位数SNR:2.56 [PC-FDK],4.76 [MA-ROOSTER],5.02 [FDK];中位数SSIM:0.18 [PC-FDK与FDK],0.31 [MA-ROOSTER与FDK])。MA-ROOSTER改善的图像质量有助于大多数重建的GTV轮廓变形和真实运动监测。

结论

基于MA-ROOSTER的4DCBCT在图像质量方面表现良好,对于(N)SCLC患者的PBS-PT治疗中的每日内部运动监测似乎很有前景。