Constantin Maria-Magdalena, Bucur Stefana, Serban Elena-Daniela, Olteanu Rodica, Bratu Ovidiu Gabriel, Constantin Traian
The Second Department of Dermatology, Colentina Clinical Hospital, 020125 Bucharest, Romania.
3rd Department, 'Carol Davila' University of Medicine and Pharmacy, 050474 Bucharest, Romania.
Exp Ther Med. 2020 Dec;20(6):184. doi: 10.3892/etm.2020.9314. Epub 2020 Oct 13.
Non-invasive bioengineering technologies are constantly being developed, as they can provide useful insights and contribute to the improvement of medical care and scientific education. The purpose of this study was to assess skin viscoelasticity using the suction chamber method in patients with allergic contact dermatitis vs. healthy subjects, before and after applying a moisturizer safety testing cream. This was a prospective controlled study over a 3-year period (March 2016-March 2019), with 81 subjects being divided in two balanced groups: Patients with allergic contact dermatitis and healthy subjects, respectively. The skin viscoelasticity was determined for all subjects with Cutometer, using the suction method, by performing a dynamic assessment of parameters before and after applying a moisturizing cream. The results indicate a decrease in the elasticity parameters in both groups, indicating an improvement of the elastic properties under the treatment. Skin capacity to return to its previous form after the deformation, i.e., pure elasticity and biological elasticity, showed overall elevated values in the group with contact dermatitis, demonstrating the efficacy of the emollient cream after applying it for 28 days (increase by 11.7 and 4.9% respectively, compared with baseline, when patients had dry, untreated skin). However, in healthy subjects, these parameters do not achieve important values, but they remain rather stable over time with a very slight improvement (6.6% after 28 days). The Cutometer is an easy to use, efficient and widely used instrument for measurements in studies that perform a quantitative assessment of the effectiveness of different formulations intended for application on the skin.
非侵入性生物工程技术不断发展,因为它们能提供有用的见解,并有助于改善医疗护理和科学教育。本研究的目的是在使用保湿剂安全测试霜之前和之后,采用吸室法评估过敏性接触性皮炎患者与健康受试者的皮肤粘弹性。这是一项为期3年(2016年3月至2019年3月)的前瞻性对照研究,81名受试者被分为两个均衡的组:分别为过敏性接触性皮炎患者和健康受试者。使用Cutometer通过吸法对所有受试者进行皮肤粘弹性测定,在涂抹保湿霜前后对参数进行动态评估。结果表明两组的弹性参数均下降,表明治疗后弹性特性有所改善。皮肤在变形后恢复到先前形状的能力,即纯弹性和生物弹性,在接触性皮炎组中总体上显示出升高的值,表明在涂抹润肤霜28天后(与患者皮肤干燥且未治疗时的基线相比,分别增加了11.7%和4.9%)润肤霜的功效。然而,在健康受试者中,这些参数没有达到显著值,但随着时间的推移它们保持相当稳定,只有非常轻微的改善(28天后为6.6%)。Cutometer是一种易于使用、高效且广泛应用于研究中的仪器,这些研究对旨在应用于皮肤的不同配方的有效性进行定量评估。