PLoS One. 2020 Oct 26;15(10):e0241296. doi: 10.1371/journal.pone.0241296. eCollection 2020.
To compare the corneal biomechanical parameters between pseudoexfoliation glaucoma (PXG), primary open-angle glaucoma (POAG) and healthy controls using Corvis ST.
A prospective, cross-sectional study was conducted which included 132 treatment-naïve eyes which underwent Corvis ST. The study cohort comprised of 44 eyes with PXG, 42 eyes with POAG and 46 healthy controls. Corneal biomechanical parameters, which included corneal velocities, length of corneal applanated surface, deformation amplitude (DA), peak distance and radius of curvature, were compared between the groups using analysis of variance models.
The 3 groups were demographically similar. The mean IOP was 15.7 ±3 mmHg in the control group, 21.3 ±5 mmHg in the POAG group and 25.8 ±7 mmHg in the PXG group (p<0.0001). Corneal pachymetry was similar across the 3 groups. Mean DA was significantly lower (p<0.0001) in the PXG group (0.86 ±0.18 mm) compared to the POAG group (0.97 ±0.14mm) and the control group (1.10 ±0.15mm). Corneal velocities were also found to be statistically significantly different between the groups. However, after adjusting for IOP, there was no difference in any of the biomechanical parameters between the 3 groups.
Corneal biomechanical parameters measured on Corvis ST are not different between eyes with PXG, POAG and healthy controls after adjusting for IOP.
使用 Corvis ST 比较原发性开角型青光眼(POAG)、假性剥脱性青光眼(PXG)和健康对照组的角膜生物力学参数。
这是一项前瞻性、横断面研究,纳入了 132 例未经治疗的接受 Corvis ST 检查的患者。研究队列包括 44 只 PXG 眼、42 只 POAG 眼和 46 只健康对照组。使用方差分析模型比较各组的角膜生物力学参数,包括角膜速度、角膜压平表面长度、变形幅度(DA)、峰值距离和曲率半径。
3 组的人口统计学特征相似。对照组的平均眼压为 15.7±3mmHg,POAG 组为 21.3±5mmHg,PXG 组为 25.8±7mmHg(p<0.0001)。3 组的角膜厚度相似。PXG 组的平均 DA 明显低于 POAG 组(0.86±0.18mm)和对照组(1.10±0.15mm)(p<0.0001)。各组的角膜速度也存在统计学差异。然而,在调整眼压后,3 组之间的任何生物力学参数均无差异。
在调整眼压后,PXG、POAG 和健康对照组的 Corvis ST 测量的角膜生物力学参数没有差异。