应用 Corvis ST 眼前节分析仪测量原发性开角型青光眼和正常眼的角膜生物力学特征。

Corneal biomechanical characteristics measured by the CorVis Scheimpflug technology in eyes with primary open-angle glaucoma and normal eyes.

机构信息

Beijing Ophthalmology and Visual Sciences Key Laboratory, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing, China.

Department of Ophthalmology, Chinese PLA General Hospital, Beijing, China.

出版信息

Acta Ophthalmol. 2016 Aug;94(5):e317-24. doi: 10.1111/aos.12672. Epub 2015 Feb 1.

Abstract

PURPOSE

To evaluate the biomechanical properties of the cornea using Corneal Visualization Scheimpflug Technology (CorVis ST, Oculus) in eyes with primary open-angle glaucoma (POAG) and normal control eyes.

METHODS

A comparative cross-sectional study that included 42 patients with POAG and 60 normal control subjects matched for intraocular pressure (IOP) and central corneal thickness (CCT). IOP was measured with a Goldmann applanation tonometer (GAT). Corneal tomography and biomechanical parameters were measured with Pentacam (Oculus) and CorVis ST, respectively. Corneal biomechanical properties were compared between groups, and the associations between corneal biomechanical parameters and ocular characteristics were evaluated. Receiver operating characteristic (ROC) curves were used to establish a cut-off value for the biomechanical parameters.

RESULTS

The following parameters of the CorVis ST showed a significant difference between eyes with POAG and normal eyes: first applanation velocity (Vin ), second applanation time (A-time2 ), peak distance (PD) and deformation amplitude (DA). In the univariate analysis, DA was negatively correlated with IOP in both groups. For all biomechanical parameters, the areas under the ROC curve were <0.80 and thus did not reach a good level of predictive accuracy for detecting POAG.

CONCLUSION

CorVis ST offers an alternative method for measuring corneal biomechanical properties. Eyes with POAG exhibit a faster Vin , longer A-time2 , lower DA and longer PD than do IOP- and CCT-matched normal control eyes. The biomechanical parameters of the CorVis ST cannot readily be used for diagnosis of POAG in the individual patient.

摘要

目的

使用角膜可视化 Scheimpflug 技术(CorVis ST,Oculus)评估原发性开角型青光眼(POAG)和正常对照眼的角膜生物力学特性。

方法

这是一项回顾性对比研究,纳入了 42 例 POAG 患者和 60 例眼压(IOP)和中央角膜厚度(CCT)匹配的正常对照者。采用 Goldmann 压平眼压计(GAT)测量 IOP。分别使用 Pentacam(Oculus)和 CorVis ST 测量角膜断层和生物力学参数。比较两组之间的角膜生物力学特性,并评估角膜生物力学参数与眼部特征之间的相关性。使用受试者工作特征(ROC)曲线为生物力学参数建立截断值。

结果

CorVis ST 的以下参数在 POAG 眼和正常眼中有显著差异:首次压平速度(Vin )、第二次压平时间(A-time2 )、峰值距离(PD)和变形幅度(DA)。在单变量分析中,两组中 DA 与 IOP 呈负相关。对于所有生物力学参数,ROC 曲线下面积均<0.80,因此未能达到良好的预测 POAG 准确性水平。

结论

CorVis ST 提供了一种测量角膜生物力学特性的替代方法。与眼压和 CCT 匹配的正常对照组相比,POAG 眼的 Vin 更快,A-time2 更长,DA 更低,PD 更长。CorVis ST 的生物力学参数不能用于个体患者 POAG 的诊断。