Bold Uyen, Tumur S, Kh Otgonchimeg, Sanjaa E
Ophthalmology Department, Orbita Eye Hospital, Ulaanbaatar, Mongolia.
Taiwan J Ophthalmol. 2019 Nov 22;10(3):217-221. doi: 10.4103/tjo.tjo_58_19. eCollection 2020 Jul-Sep.
The purpose is to present the objective of the Sanjaa-Uyen-Tumur (SUT) ophthalmometer and compare its measurements with the standard Hertel exophthalmometer.
Eyeball position of 88 healthy eyes of the patients was measured using both the SUT ophthalmometer and Hertel exophthalmometer. Both methods were performed in one session by the same experienced ophthalmologist. Data were analyzed using Bland-Altman method.
Mean age of the participants was 41.11 ± 14.08, further, 59% (26) were male and 41% (18) were female, respectively. The mean difference was 0.1420455 and the standard deviation of the difference was 0.9221067, 95% confidence interval, respectively. Based on the Bland-Altman analysis, lower limit of agreement in our study was from -2.284538 to -1.614211 and upper limit of agreement was from 1.33012 to 2.000447. According to the results of our studies, there is no relationship between the difference and the level of measurement in either plot.
This SUT ophthalmometer can be used same as the Hertel exophthalmometer in ophthalmology practice as can be accurate, affordable, and objective.
介绍桑贾-乌延-图穆尔(SUT)眼球突出计的目的,并将其测量结果与标准的赫特尔眼球突出计进行比较。
使用SUT眼球突出计和赫特尔眼球突出计测量患者88只健康眼睛的眼球位置。两种方法均由同一位经验丰富的眼科医生在一次检查中进行。采用布兰德-奥特曼方法分析数据。
参与者的平均年龄为41.11±14.08岁,此外,男性占59%(26人),女性占41%(18人)。平均差值为0.1420455,差值的标准差分别为0.9221067,95%置信区间。根据布兰德-奥特曼分析,本研究中的一致性下限为-2.284538至-1.614211,一致性上限为1.33012至2.000447。根据我们的研究结果,在任何一个图表中,差值与测量水平之间均无关系。
这种SUT眼球突出计在眼科实践中可与赫特尔眼球突出计一样使用,因为它准确、经济且客观。