Immunoassays-Immunosensors Lab, Institute of Nuclear & Radiological Sciences & Technology, Safety & Energy, National Center for Scientific Research "Demokritos", 15310 Aghia Paraskevi, Greece.
Analytical Chemistry Lab, Department of Chemistry, National and Kapodistrian University of Athens, Panepistimiopolis, 15771 Zografou, Greece.
Biosensors (Basel). 2020 Oct 25;10(11):154. doi: 10.3390/bios10110154.
Deoxynivalenol (DON) is a mycotoxin produced by certain species and found in a high percentage of wheat and maize grains cultured worldwide. Although not so toxic as other mycotoxins, it exhibits both chronic and acute toxicity, and therefore methods for its fast and accurate on-site determination are highly desirable. In the current work, we employ an optical immunosensor based on White Light Reflectance Spectroscopy (WLRS) for the fast and sensitive immunochemical label-free determination of DON in wheat and maize samples. The assay is completed in 12 min and has a quantification limit of 2.5 ng/mL in buffer corresponding to 125 μg/kg in whole grain which is lower than the maximum allowable concentrations set by the regulatory authorities for grains intended for human consumption. Several extraction protocols have been compared, and the highest recovery (>90%) was achieved employing distilled water. In addition, identical calibration curves were received in buffer and wheat/maize extraction matrix providing the ability to analyze the grain samples using calibrators in buffer. Recoveries of DON from spiked wheat and maize grain samples ranged from 92.0(±4.0) to 105(±4.0)%. The analytical performance of the WLRS immunosensor, combined with the short analysis time and instrument portability, supports its potential for on-site determinations.
脱氧雪腐镰刀菌烯醇(DON)是某些真菌产生的一种真菌毒素,在世界范围内培养的小麦和玉米谷物中大量存在。虽然不如其他真菌毒素那么有毒,但它具有慢性和急性毒性,因此非常需要快速准确的现场测定方法。在当前的工作中,我们采用基于白光反射光谱(WLRS)的光学免疫传感器,用于快速、灵敏地免疫化学无标记测定小麦和玉米样品中的 DON。该测定在 12 分钟内完成,在缓冲液中的定量限为 2.5ng/mL,相当于全谷物中的 125μg/kg,低于监管机构为人类食用谷物设定的最大允许浓度。比较了几种提取方案,采用蒸馏水可获得最高回收率(>90%)。此外,在缓冲液和小麦/玉米提取基质中均获得了相同的校准曲线,这使得可以使用缓冲液中的校准器来分析谷物样品。从添加了 DON 的小麦和玉米谷物样品中的回收率范围为 92.0(±4.0)至 105(±4.0)%。WLRS 免疫传感器的分析性能,结合短的分析时间和仪器便携性,支持其用于现场测定的潜力。