Kang Jung Youb, Song Ju Hwan, Lee Sang Joon
Department of Ophthalmology, Kosin University College of Medicine, Busan, South Korea.
The Bom Eye Clinic, Busan, South Korea.
Clin Ophthalmol. 2020 Oct 14;14:3343-3349. doi: 10.2147/OPTH.S277305. eCollection 2020.
To investigate optic opacification of hydrophobic acrylic intraocular lens (IOL) depending on hydration and temperature conditions.
IOL (Acrysof MA60BM Alcon, Fort Worth, Texas, USA) extracted from a 55-year-old male who underwent binocular cataract 11 years ago were used for the analysis. On slit lamp examination, the binocular IOL optics showed homogenous opacification. After extraction of the IOLs, the IOL opacification disappeared during dry storage at 4°C. To investigate the changes in IOL opacification according to temperature and hydration, an extracted IOL was stored in either dry conditions or normal saline at 4°C, room temperature, or 37°C for 24 hours, and then the degree of haziness was examined. To investigate the impact of an opaque optic on visual function, light transmission was performed using an ultraviolet-visible spectrophotometer (U-3000, Hitachi High-Technologies Corp., Tokyo, Japan).
The extracted IOL optics at 4°C, room temperature, and 37°C were transparent at dry conditions. When the dried IOL was immersed in normal saline at room temperature and 37°C, opacification appeared. However, when the dried IOL was immersed in normal saline at 4°C, opacification of the IOL did not appear. When compared with the control, light transmission of the extracted IOLs from the right and left eyes stored in 37°C normal saline decreased by 4.7% at 453 nm and by 5.1% at 482 nm, respectively.
In acrylic hydrophobic IOLs, optic opacification can occur depending on temperature and hydration conditions.
研究疏水丙烯酸人工晶状体(IOL)的光学混浊与水化和温度条件的关系。
分析11年前接受双眼白内障手术的55岁男性患者所使用的IOL(美国德克萨斯州沃思堡爱尔康公司的Acrysof MA60BM)。裂隙灯检查显示双眼IOL光学部呈均匀混浊。IOL取出后,在4°C干燥保存期间IOL混浊消失。为研究IOL混浊随温度和水化的变化,将取出的IOL分别在4°C、室温或37°C的干燥条件或生理盐水中保存24小时,然后检查混浊程度。为研究不透明光学部对视觉功能的影响,使用紫外可见分光光度计(日本东京日立高新技术公司的U - 3000)进行透光率检测。
在4°C、室温及37°C条件下,取出的IOL光学部在干燥状态下是透明的。当干燥的IOL浸入室温及37°C的生理盐水中时,出现混浊。然而,当干燥的IOL浸入4°C的生理盐水中时,IOL未出现混浊。与对照组相比,保存在37°C生理盐水中的左右眼取出的IOL在453 nm处的透光率分别下降了4.7%,在482 nm处下降了5.1%。
对于丙烯酸疏水IOL,光学混浊可能会根据温度和水化条件而发生。