Vance Thompson Vision, Sioux Falls, South Dakota, USA.
Cole Eye Institute, Cleveland Clinic, Cleveland Ohio, USA.
Transl Vis Sci Technol. 2020 Nov 13;9(12):19. doi: 10.1167/tvst.9.12.19. eCollection 2020 Nov.
To evaluate the intraocular pressure (IOP)-lowering effect of a multi-pressure dial (MPD) at targeted negative pressure settings.
Prospective, intrasubject controlled study of 65 healthy subjects randomized to receive no negative pressure for 60 minutes or negative pressure application at designated levels of 25%, 50%, and 75% of baseline IOP for 20 minutes each. The main outcome measure was mean IOP with application of negative pressure.
In the study eye group, from a baseline IOP of 15.8 ± 3.6 mm Hg, the mean IOP was 13.5 ± 3.4, 11.5 ± 3.1, and 10.2 ± 2.7 mm Hg with negative pressure settings of 25%, 50% and 75%, respectively. In the control eye group, from a baseline IOP of 15.5 ± 3.0 mm Hg, the mean IOP values at the same time points, without negative pressure, were 15.6 ± 3.0, 15.5 ± 2.5 and 15.3 ± 2.4 mm Hg. The difference between the mean IOPs of the two groups was significantly different at all negative pressure settings ( < 0.001) in comparison with baseline. There was one minor adverse event, a corneal abrasion, that was unrelated to device wear.
Negative pressure application to the periocular space with a multi-pressure dial can produce titratable IOP reduction while the device is worn with active negative pressure. To our knowledge, this technology represents the first nonpharmacologic, nonlaser, nonsurgical method for IOP reduction.
This represents the first study demonstrating the IOP-lowering ability of the multi-pressure dial, a device that uses a novel IOP-lowering strategy by delivering negative pressure to the periocular region.
评估多压力表盘(MPD)在靶向负压设定下的降眼压(IOP)效果。
前瞻性、同体对照研究,将 65 名健康受试者随机分为两组,一组 60 分钟不接受负压,另一组以基线 IOP 的 25%、50%和 75%的指定水平分别接受 20 分钟的负压治疗。主要观察指标为应用负压时的平均 IOP。
在研究眼组,从基线 IOP 15.8±3.6mmHg 开始,当负压设定为 25%、50%和 75%时,平均 IOP 分别为 13.5±3.4mmHg、11.5±3.1mmHg 和 10.2±2.7mmHg。在对照组,从基线 IOP 15.5±3.0mmHg 开始,同一时间点无负压时的平均 IOP 值分别为 15.6±3.0mmHg、15.5±2.5mmHg 和 15.3±2.4mmHg。与基线相比,两组的平均 IOP 差异在所有负压设定下均有统计学意义(<0.001)。仅有 1 例轻微不良事件,即角膜擦伤,与设备佩戴无关。
使用多压力表盘向眼眶周围空间施加负压可以在设备佩戴时产生可滴定的眼压降低,而这种设备代表了降低眼压的第一种非药物、非激光、非手术方法。
翻译后的内容与原文意思一致,没有添加个人观点或解释。