WaveOne Gold, ProTaper Gold, Reciproc Blue, ProTaper Next, WaveOne and ProTaper files were selected to compare the phase transformation behaviors and mechanical properties of nickel-titanium (NiTi) rotary files after gold heat and blue heat treatments. The reverse transformation finishing point temperatures of WaveOne Gold, ProTaper Gold, ProTaper Next and WaveOne were higher than those of the other two instruments investigated. At a deflection of 0.5 mm, the loads were significantly varied except for ProTaper Next and ProTaper Gold. At a deflection of 3.0 mm, the loads of Reciproc Blue and WaveOne Gold were significantly varied compared to WaveOne and ProTaper. Cycles to failure were reduced in the order of WaveOne Gold, Reciproc Blue, ProTaper Gold, ProTaper Next, WaveOne and ProTaper. NiTi instruments after gold heat and blue heat treatments exhibited significantly higher bending properties and cyclic fatigue resistances, representing an improved performance over traditional and M-wire instruments.
选择WaveOne Gold、ProTaper Gold、Reciproc Blue、ProTaper Next、WaveOne和ProTaper锉来比较镍钛(NiTi)旋转锉在金热和蓝热治疗后的相变行为和力学性能。WaveOne Gold、ProTaper Gold、ProTaper Next和WaveOne的逆转变终点温度高于所研究的其他两种器械。在0.5毫米的挠度下,除ProTaper Next和ProTaper Gold外,载荷有显著变化。在3.0毫米的挠度下,与WaveOne和ProTaper相比,Reciproc Blue和WaveOne Gold的载荷有显著变化。失效循环次数按WaveOne Gold、Reciproc Blue、ProTaper Gold、ProTaper Next、WaveOne和ProTaper的顺序减少。经过金热和蓝热治疗的NiTi器械表现出显著更高的弯曲性能和循环疲劳抗性,代表了比传统和M型丝器械更好的性能。