在根管预备前后,WaveOne、WaveOne Gold、Reciproc 和 Reciproc blue 的表面纳米级轮廓。

Surface nanoscale profile of WaveOne, WaveOne Gold, Reciproc, and Reciproc blue, before and after root canal preparation.

机构信息

College of Dentistry, Taibah University, P O Box 344, Madinah Al Munawwarah, Saudi Arabia.

College of Engineering, Taibah University, Madinah Al Munawwarah, Saudi Arabia.

出版信息

Odontology. 2019 Oct;107(4):500-506. doi: 10.1007/s10266-019-00424-8. Epub 2019 Mar 29.

Abstract

This study aimed to compare the surface topographies and nanoscale profiles of intact and used WaveOne, WaveOne Gold, Reciproc, and Reciproc blue nickel-titanium (NiTi) instruments. Sixteen NiTi instruments were evaluated; four from each one of WaveOne (25/0.08), WaveOne Gold (25/0.07), Reciproc (25/0.08), and Reciproc blue (25/0.08) files. Each instrument was used to prepare four curved canals. A three-dimensional nanoscale surface profiler was used to assess instrument surfaces before and after root canal instrumentation. The data were analyzed with one-way analysis of variance using SPSS 21.0 software and set significance at P < 0.05. There were significant differences in surface topography between NiTi instruments. Intact WaveOne and WaveOne Gold instruments had the highest values, followed by the Reciproc and Reciproc blue instruments (P < 0.05). Used WaveOne and WaveOne Gold instruments had the highest values, followed by Reciproc and Reciproc blue instruments (P < 0.05). Some indications of microcracks in qualitatively intact Reciproc blue files were observed, and minor surface cracks in used WaveOne Gold instruments were also observed. Used WaveOne, Reciproc, and Reciproc blue instruments exhibited cracks and microcavities. WaveOne Gold and Reciproc blue instruments had significantly higher levels of surface distortion after root canal instrumentation, and WaveOne exhibited the lowest level of surface deformation. Each of the Reciprocating instruments was safely used to prepare four canals without incident. The three-dimensional stylus profiler provided valuable assessments of the surface topographies of NiTi instruments.

摘要

本研究旨在比较未使用和使用后的 WaveOne、WaveOne Gold、Reciproc 和 Reciproc blue 镍钛(NiTi)器械的表面形貌和纳米级形貌。评估了 16 个 NiTi 器械,每个品牌各 4 个,分别为 WaveOne(25/0.08)、WaveOne Gold(25/0.07)、Reciproc(25/0.08)和 Reciproc blue(25/0.08)锉。每个器械都用于预备 4 个弯曲根管。使用三维纳米级表面轮廓仪在根管预备前后评估器械表面。使用 SPSS 21.0 软件进行单向方差分析对数据进行分析,并设置 P<0.05 为显著性水平。NiTi 器械的表面形貌存在显著差异。未使用的 WaveOne 和 WaveOne Gold 器械的表面粗糙度值最高,其次是 Reciproc 和 Reciproc blue 器械(P<0.05)。使用后的 WaveOne 和 WaveOne Gold 器械的表面粗糙度值最高,其次是 Reciproc 和 Reciproc blue 器械(P<0.05)。定性上完整的 Reciproc blue 锉上观察到一些微裂纹的迹象,使用后的 WaveOne Gold 器械上也观察到了较小的表面裂纹。使用后的 WaveOne、Reciproc 和 Reciproc blue 器械出现了裂纹和微腔。WaveOne Gold 和 Reciproc blue 器械在根管预备后表面变形程度显著更高,而 WaveOne 器械的表面变形程度最低。每个往复式器械都安全地用于预备 4 个根管,没有发生事故。三维触针轮廓仪为 NiTi 器械的表面形貌提供了有价值的评估。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验