Turk J Ophthalmol. 2020 Oct 30;50(5):264-270. doi: 10.4274/tjo.galenos.2020.18488.
To investigate the agreement between optical coherence tomography (OCT) and OCT-based angiography (OCT-A) in estimating retinal nerve fiber layer thickness (RNFLT) and evaluate the associations between peripapillary vessel density (VD) and RNFLT measurements obtained with both devices.
The AngioVue (Optovue Inc., Fremont, CA, USA) and Spectralis (Heidelberg Engineering, Heidelberg, Germany) images of 325 patients were screened retrospectively. RNFLT values were recorded using both devices. The intraclass correlation coefficient (ICC) and Bland-Altman plots were obtained to investigate the agreement between the devices. Age- and intraocular pressure-corrected associations between VD and RNFLT measured by the two devices were analyzed using linear regression models.
ICC revealed excellent agreement for global, superior, inferior, and temporal RNFLT and good agreement for the nasal quadrant (ICC=0.895, 0.936, 0.923, 0.887, and 0.614, respectively). The Bland-Altman plots showed poor agreement for all measurements with a large span of limits of agreement and significant proportional bias (p<0.05). VD was found to be strongly associated with the RNFLT measurements of both devices (p<0.001).
The disagreement between the devices should be considered in clinical practice, and the data should not be used interchangeably. The association of the peripapillary VD with RNFLT using both devices indicated that RNFLT assessed by the AngioVue could be used in glaucoma management along with VD.
研究光学相干断层扫描(OCT)和基于 OCT 的血管造影(OCT-A)在评估视网膜神经纤维层厚度(RNFLT)方面的一致性,并评估两种设备获得的视盘周围血管密度(VD)和 RNFLT 测量值之间的相关性。
回顾性筛选了 325 名患者的 AngioVue(Optovue Inc.,加利福尼亚州弗里蒙特)和 Spectralis(Heidelberg Engineering,德国海德堡)图像。使用两种设备记录 RNFLT 值。使用组内相关系数(ICC)和 Bland-Altman 图来研究设备之间的一致性。使用线性回归模型分析两种设备测量的 VD 与 RNFLT 之间的年龄和眼内压校正相关性。
ICC 显示,全局、上、下和颞部 RNFLT 的一致性极好,鼻象限的一致性良好(ICC=0.895、0.936、0.923、0.887 和 0.614)。Bland-Altman 图显示,所有测量值的一致性都很差,一致性界限范围较大,存在显著的比例偏差(p<0.05)。VD 与两种设备的 RNFLT 测量值均呈强相关(p<0.001)。
在临床实践中应考虑设备之间的差异,数据不能互换使用。两种设备的视盘周围 VD 与 RNFLT 的相关性表明,AngioVue 评估的 RNFLT 可以与 VD 一起用于青光眼管理。